FDA Adverse Event Malfunction Summary report: N

BD¿ PRN ADAPTER

MDR report key: 15225771 · Received August 15, 2022

Report

Report Number
3014704491-2022-00353
Event Type
Malfunction
Date Received
August 15, 2022
Date of Event
July 13, 2022
Report Date
August 5, 2022
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FPA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED, A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1173181. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. BASED ON THE DESCRIPTION OF THE EVENT AND PREVIOUS INVESTIGATIONS INTO THIS PRODUCT FAMILY OUR ENGINEER SHAVE DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR THIS EVENT IS THE PRESENCE OF SILICONE RESIDUE ON THE SLEEVE STOPPER. SILICONE IS USED IN THE MANUFACTURE OF THIS DEVICE AND CAN REDUCE THE FRICTION IMPARTED ON THE SURFACE OF THE SLEEVE STOPPER. TO MONITOR OCCURRENCE SUCH AS THESE OUR TEAM IMPLEMENTS A PULL TEST TO CONFIRM THAT THE SLEEVE STOPPER IS PROPERLY FIXED INTO ITS POSITION; WE HAVE INCREASED THE FORCE REQUIREMENTS FOR THIS TEST AS WELL AS UPDATED OUR FINISHED GOODS INSPECTION LIST AND ADJUST TEMPERATURE SETTINGS ON THE HEAT TUNNELS TO INCREASE THE GRIP OF THE SLEEVE STOPPER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD¿ PRN ADAPTER STOPPER FELL OFF DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: THE INFUSION NEEDLE WAS CONNECTED TO THE PRN AND THEN THE TUBE WAS RINSED, THE YELLOW STOPPER OF THE PRN FELL OFF, AND THE HEPARIN CAP WAS REPLACED AND THE INFUSION SET WAS RE-ADMINISTERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1941895 BD¿ PRN ADAPTER INTRAVASCULAR ADMINISTRATION SET FPA BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 1173181

Patients

Seq Age Sex Outcome Treatment
1 Unknown