FDA Adverse Event Malfunction Summary report: N

SUPERDIMENSION BRONCHUS SYSTEM (SDBS)

MDR report key: 1522238 · Received October 27, 2009

Report

Report Number
3004659744-2009-00031
Event Type
Malfunction
Date Received
October 27, 2009
Date of Event
September 30, 2009
Report Date
October 28, 2009
Manufacturer
SUPERDIMENSION, LTD.
Product Code
JAK
PMA / PMN Number
K042438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN ON-SITE VISIT WAS PERFORMED ON OCTOBER 19 AND 20, 2009. DURING THE ON-SITE VISIT THE SYSTEM PERFORMED WITHIN SPEC. IN ADDITION, 2 LOCATABLE GUIDES (LG'S) WERE RETURNED TO SUPERDIMENSION FOR A PRECAUTIONARY ANALYSIS. BOTH LG'S WERE TESTED AND WERE ALSO WITHIN SPECS. THE ERROR MESSAGE INDICATES THE PHYSICIAN WAS NOT MARKING THE SAME REGISTRATION POINTS IN THE PT AS WERE MARKED DURING PLANNING. THE PROCEDURE IS TYPICALLY PERFORMED UNDER LOCAL ANESTHESIA. HOWEVER, BASED ON THE ERROR MESSAGES PROVIDED BY THE SYSTEM, THE PROCEDURE WAS DISCONTINUED, BY THE PHYSICIAN, WITH THE PT UNDER GENERAL ANESTHESIA. IN AN ABUNDANCE OF CAUTION, THIS EVENT IS BEING REPORTED DUE TO THE ADD'L POTENTIAL RISK ASSOCIATED WITH MULTIPLE EXPOSURE TO GENERAL ANESTHESIA. THERE WAS NO INJURY TO THE PT REPORTED.

Description of Event or Problem · 1

THE PHYSICIAN ATTEMPTED TO REGISTER THE PT MULTIPLE TIMES AND ALWAYS REC'D THE ERROR MESSAGE STATING "SOME OF THE REGISTRATION POINTS YOU HAVE ACQUIRED ARE MISMATCHED WITH THE POINTS PREVIOUSLY MARKED IN THE PLANNING PHASE. PLEASE VERIFY THAT THE POINTS HAVE BEEN ACQUIRED CORRECTLY, AND RE-ACQUIRE ALL POINTS IF POSSIBLE." THE SITE SWAPPED THE LOCATABLE GUIDE, BUT THIS DID NOT HELP, THEY CONTINUED TO RECEIVE THE ERROR MESSAGE. THEREFORE, THE PROCEDURE WAS CANCELLED WITH THE PT UNDER GENERAL ANESTHESIA. THERE WAS NO HARM OR INJURY TO THE PT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUPERDIMENSION BRONCHUS SYSTEM (SDBS) COMPUTED TOMOGRAPHY X-RAY SYSTEM JAK SUPERDIMENSION, LTD.

Patients

Seq Age Sex Outcome Treatment
1