BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBE
Report
- Report Number
- 9617032-2022-00762
- Event Type
- Malfunction
- Date Received
- August 12, 2022
- Date of Event
- July 20, 2022
- Report Date
- September 12, 2022
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1309955, MEDICAL DEVICE EXPIRATION DATE: 2023-04-30, DEVICE MANUFACTURE DATE: 2021-11-05. MEDICAL DEVICE LOT #: 2094463, MEDICAL DEVICE EXPIRATION DATE: 2023-09-30, DEVICE MANUFACTURE DATE: 2022-04-04. MEDICAL DEVICE LOT #: 1320310, MEDICAL DEVICE EXPIRATION DATE: 2023-05-31, DEVICE MANUFACTURE DATE: 2021-11-16. MEDICAL DEVICE LOT #: 2040913, MEDICAL DEVICE EXPIRATION DATE: 2023-08-31, DEVICE MANUFACTURE DATE: 2022-02-09. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. THEREFORE, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES RELATING TO ERRONEOUS RESULTS WERE OBSERVED. THERE WERE NO DIFFICULTIES ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES APPEARED TO EXHIBIT PROPER FILL. BD WAS UNABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE, ERRONEOUS RESULTS, BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE COMPLAINT LOT SAMPLES. REPLICATES OF BOTH RETAIN AND CONTROL SAMPLES WERE ACCEPTABLE IN TERMS OF BOTH PRECISION AND ACCURACY. ALL VISUAL OBSERVATIONS OF BOTH RETAIN AND CONTROL SAMPLES TESTED DEMONSTRATED CLINICALLY ACCEPTABLE PERFORMANCE. ALL TUBES PERFORMED AS EXPECTED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE ERRONEOUS RESULTS. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. FACTORS THAT MAY CONTRIBUTE TO ERRONEOUS RESULTS WERE EVALUATED THROUGH A CLINICAL STUDY TO VERIFY THE DESIGN OF THE DEVICE MET IT¿S INTENDED USE. THE RESULT OF THE STUDY SHOWED THAT THE DEVICE PERFORMED AS EXPECTED AND WE WERE UNABLE TO DETERMINE ANY EXTERNAL CONTRIBUTOR TO THIS REPORTED ISSUE.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBE THERE WAS ERRONEOUS RESULTS. THIS EVENT OCCURRED 400 TIMES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "WE PREDOMINANTLY USE VACUTAINER SST II ADVANCE BLOOD TUBES. GENERAL PRACTITIONERS HAVE REPORTED FALSELY ELEVATED PHOSPHATE RESULTS. THE ISSUE APPEARS TO BE PREVALENT. PRELIMINARY INVESTIGATIONS HAVE RULED OUT AN ANALYTICAL ISSUE. PHOSPHATE RESULTS ARE REPRODUCIBLE USING DIFFERENT MEASURING SYSTEMS. INSTRUMENT HARDWARE CHECKS USED TO DETECT REAGENT CARRIER FROM ON-BOARD REAGENTS CONTAINING PHOSPHATE BUFFERS HAVE RETURNED FAVORABLY. ELEVATED PHOSPHATE RESULTS ARE NOT ASSOCIATED WITH ELEVATED POTASSIUM RESULTS, AN INDICATION OF DELAYED SEPARATION. THE ELEVATION OF PHOSPHATE IS CONSISTENT ACROSS PATIENTS APPROXIMATELY +1MMOL/L. SO FAR AFFECTED PATIENTS HAVE BEEN ISOLATED TO THE FOLLOWING LOT NUMBERS OF TUBES; (B)(6)."
IT WAS REPORTED WHEN USING THE BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBE THERE WAS ERRONEOUS RESULTS. THIS EVENT OCCURRED 400 TIMES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "WE PREDOMINANTLY USE VACUTAINER SST II ADVANCE BLOOD TUBES. GENERAL PRACTITIONERS HAVE REPORTED FALSELY ELEVATED PHOSPHATE RESULTS. THE ISSUE APPEARS TO BE PREVALENT. PRELIMINARY INVESTIGATIONS HAVE RULED OUT AN ANALYTICAL ISSUE. PHOSPHATE RESULTS ARE REPRODUCIBLE USING DIFFERENT MEASURING SYSTEMS. INSTRUMENT HARDWARE CHECKS USED TO DETECT REAGENT CARRIER FROM ON-BOARD REAGENTS CONTAINING PHOSPHATE BUFFERS HAVE RETURNED FAVORABLY. ELEVATED PHOSPHATE RESULTS ARE NOT ASSOCIATED WITH ELEVATED POTASSIUM RESULTS, AN INDICATION OF DELAYED SEPARATION. THE ELEVATION OF PHOSPHATE IS CONSISTENT ACROSS PATIENTS APPROXIMATELY +1MMOL/L. SO FAR AFFECTED PATIENTS HAVE BEEN ISOLATED TO THE FOLLOWING LOT NUMBERS OF TUBES; 1309955, 2094463, 1320310, 2040913."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2848057 | BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBE | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |