FDA Adverse Event Malfunction Summary report: N

QUICK SET PARADIGM

MDR report key: 1522109 · Received November 2, 2009

Report

Report Number
8021545-2009-00036
Event Type
Malfunction
Date Received
November 2, 2009
Date of Event
June 30, 2009
Report Date
October 29, 2009
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
PMA / PMN Number
K991759
Removal / Correction Number
Z-1705-2009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: ONE USED DEVICE WAS RETURNED FOR EVALUATION. USED DEVICES: THE TUBING WAS VISUALLY INSPECTED FOR ANY TEARS OR CRACKS ON THE TUBING IT SELF AND ON THE ATTACHED CONNECTOR PARTS. FURTHERMORE A FLOW TEST AND A P-CAP CONNECTOR VENT TEST WERE PERFORMED. THE MEMBRANE IN THE P-CAP CONNECTOR WAS HUMID AND THE SAMPLE ALSO FAILED THE P-CAP CONNECTOR VENT TEST. FURTHERMORE THE USED SAMPLE FAILED FLOW TEST OF THE TUBING. REFERENCE SAMPLES: THE REFERENCE MATERIAL WAS TESTED AND ALL SAMPLES WERE FOUND TO BE WITHIN SPECIFICATIONS. A REVIEW OF THE RELEVANT DEVICE HISTORY RECORDS AND A SEARCH FOR RELEVANT DEVIATIONS OR SIMILAR COMPLAINTS RELATED TO THE SPECIFIC DEVICE RESULTED IN NO SIMILAR COMPLAINTS FOR THE INVOLVED LOT NUMBER 8201039. NO RELEVANT DEVIATIONS DURING MANUFACTURING WERE RECORDED.

Description of Event or Problem · 1

PATIENT STATES THAT INSULIN CONTINUED TO EXIT TUBING AFTER PUMP MOTOR HAD SLOWED DOWN. PATIENT STATES THERE IS NO GAP BETWEEN PUMP PISTON AND RESERVOIR. UNOMEDICAL RECEIVED THE COMPLAINT THROUGH (B)(4), THE DISTRIBUTOR OF THE INFUSION SET. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUICK SET PARADIGM FPA UNOMEDICAL A/S MMT-398 8201039

Patients

Seq Age Sex Outcome Treatment
1 NA