FDA Adverse Event Death Summary report: N

RESUSCITAIRE INFANT RADIANT WARMER

MDR report key: 1521813 · Received November 3, 2009

Report

Report Number
2510954-2009-00008
Event Type
Death
Date Received
November 3, 2009
Date of Event
September 24, 2009
Report Date
November 3, 2009
Manufacturer
DRAGER MEDICAL SYSTEMS, INC. (PNC)
Product Code
FMT
PMA / PMN Number
K003335
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER INDICATED THE GAS CONFIGURATION AT THE USER FACILITY UTILIZES ONE GAS SOURCE TO BE WALL (O2) AND THE OTHER TO BE RESERVE TANK SOURCE (AIR). A DRAEGER SERVICE REP WAS DISPATCHED AND OBSERVED A THIRD PARTY DISPOSABLE PT CIRCUIT WAS USED INSTEAD OF THE MANUFACTURER PART. THE BLENDER ALARMED AS DESIGNED TO INDICATE LOSS OF GAS PRESSURE SOURCE. THIS OCCURS WHEN THE PRESSURE DIFFERENTIAL BETWEEN THE GAS SOURCES EXCEEDS 207 KPA (30 PSI) +/- 14 KPA (2 PSI). THE INSTRUCTIONS FOR USE INCLUDES A WARNING STATING: "IF EITHER THE AIR OR OXYGEN GAS SOURCE IS REDUCED, CREATING A PRESSURE DIFFERENTIAL OF 207 KPA (30 PSI), THE BLENDER ALARM SOUNDS. THIS CONDITION CAN ALTER THE FIO2 AND FLOW OUTPUT FROM THE MICROBLENDER. INFANT INJURY COULD OCCUR." THE USER FACILITY'S RESERVE TANK SOURCE (AIR) CAPACITY DEPLETED, THEREBY CAUSING THE ALARM CONDITION. DEVICE OPERATION WAS CHECKED BY A DRAEGER SERVICE REP AND FOUND TO BE FUNCTIONAL. BASED ON THE INFO SUPPLIED TO DATE, THERE IS NO INDICATION OF A DEVICE MALFUNCTION.

Description of Event or Problem · 1

IN THE PROCESS OF A DIFFICULT RESUSCITATION OF PT, THE USER FACILITY'S RESERVE TANK (AIR) CAPACITY RAN OUT CAUSING INFLATION PRESSURE TO DROP TO A LEVEL WHERE INSUFFICIENT PRESSURE WAS AVAILABLE TO CONTINUE RESUSCITATION. THE BLENDER ALARMED TO ALERT THE USER TO THE CONDITION; HOWEVER, THEY WERE UNABLE TO CORRECT THE SITUATION AND THE PT EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESUSCITAIRE INFANT RADIANT WARMER INFANT RADIANT WARMER FMT DRAGER MEDICAL SYSTEMS, INC. (PNC) RW82 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Death