RESUSCITAIRE INFANT RADIANT WARMER
Report
- Report Number
- 2510954-2009-00008
- Event Type
- Death
- Date Received
- November 3, 2009
- Date of Event
- September 24, 2009
- Report Date
- November 3, 2009
- Manufacturer
- DRAGER MEDICAL SYSTEMS, INC. (PNC)
- Product Code
- FMT
- PMA / PMN Number
- K003335
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER INDICATED THE GAS CONFIGURATION AT THE USER FACILITY UTILIZES ONE GAS SOURCE TO BE WALL (O2) AND THE OTHER TO BE RESERVE TANK SOURCE (AIR). A DRAEGER SERVICE REP WAS DISPATCHED AND OBSERVED A THIRD PARTY DISPOSABLE PT CIRCUIT WAS USED INSTEAD OF THE MANUFACTURER PART. THE BLENDER ALARMED AS DESIGNED TO INDICATE LOSS OF GAS PRESSURE SOURCE. THIS OCCURS WHEN THE PRESSURE DIFFERENTIAL BETWEEN THE GAS SOURCES EXCEEDS 207 KPA (30 PSI) +/- 14 KPA (2 PSI). THE INSTRUCTIONS FOR USE INCLUDES A WARNING STATING: "IF EITHER THE AIR OR OXYGEN GAS SOURCE IS REDUCED, CREATING A PRESSURE DIFFERENTIAL OF 207 KPA (30 PSI), THE BLENDER ALARM SOUNDS. THIS CONDITION CAN ALTER THE FIO2 AND FLOW OUTPUT FROM THE MICROBLENDER. INFANT INJURY COULD OCCUR." THE USER FACILITY'S RESERVE TANK SOURCE (AIR) CAPACITY DEPLETED, THEREBY CAUSING THE ALARM CONDITION. DEVICE OPERATION WAS CHECKED BY A DRAEGER SERVICE REP AND FOUND TO BE FUNCTIONAL. BASED ON THE INFO SUPPLIED TO DATE, THERE IS NO INDICATION OF A DEVICE MALFUNCTION.
IN THE PROCESS OF A DIFFICULT RESUSCITATION OF PT, THE USER FACILITY'S RESERVE TANK (AIR) CAPACITY RAN OUT CAUSING INFLATION PRESSURE TO DROP TO A LEVEL WHERE INSUFFICIENT PRESSURE WAS AVAILABLE TO CONTINUE RESUSCITATION. THE BLENDER ALARMED TO ALERT THE USER TO THE CONDITION; HOWEVER, THEY WERE UNABLE TO CORRECT THE SITUATION AND THE PT EXPIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESUSCITAIRE INFANT RADIANT WARMER | INFANT RADIANT WARMER | FMT | DRAGER MEDICAL SYSTEMS, INC. (PNC) | RW82 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Death |