FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 ANTIGEN SELF-TEST

MDR report key: 15213664 · Received August 11, 2022

Report

Report Number
1221359-2022-04262
Event Type
Malfunction
Date Received
August 11, 2022
Date of Event
July 31, 2022
Report Date
August 11, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 182411 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 195-160 / LOT 182411 AND DEVICE PART NUMBER 195-430WL / LOT 180605. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 182411 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2022 WITH A KITTED NASAL SWAB SAMPLE. CONFIRMATION PCR TESTING (PLATFORM UNKNOWN) WAS PERFORMED (B)(6) 2022 WHICH GENERATED A POSITIVE RESULT. THE CONSUMER REPORTED PERFORMING A SECOND HOME COVID-19 TEST ON (B)(6) 2022 WHICH GENERATED A POSITIVE RESULT. NO ADDITIONAL INFORMATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1163954 BINAXNOW COVID-19 ANTIGEN SELF-TEST CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 182411 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 51 YR Female