BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2022-00734
- Event Type
- Malfunction
- Date Received
- August 10, 2022
- Date of Event
- July 13, 2022
- Report Date
- August 18, 2022
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- PMA / PMN Number
- UNKNOWN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION SUMMARY MATERIAL #: 367377, LOT/BATCH #: 2032146. BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR POOR BARRIER WAS NOT OBSERVED. ADDITIONALLY, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES RELATING TO POOR BARRIER WERE OBSERVED. NO DIFFICULTIES WERE ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES APPEARED TO EXHIBIT PROPER FILL. BD WAS UNABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE (POOR BARRIER) BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE COMPLAINT LOT SAMPLES. ALL SAMPLES (RETAIN AND CONTROL LOTS) DEMONSTRATED CLINICALLY ACCEPTABLE PERFORMANCE FOR ALL VISUAL OBSERVATIONS EVALUATED. ALL TUBES PERFORMED AS EXPECTED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE POOR BARRIER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. FACTORS THAT MAY CONTRIBUTE TO POOR BARRIER WERE EVALUATED THROUGH A CLINICAL STUDY TO VERIFY THE DESIGN OF THE DEVICE MET IT¿S INTENDED USE. THE RESULT OF THE STUDY SHOWED THAT THE DEVICE PERFORMED AS EXPECTED AND WE WERE UNABLE TO DETERMINE ANY EXTERNAL CONTRIBUTOR TO THIS REPORTED ISSUE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES HAVE POOR SEPARATION BARRIER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "PLASMA STILL BELOW GEL AFTER CENTRIFUGATION, CENTRIFUGE SPEED 3500RPM FOR 1 MINUTES.CENTRIFUGE USED HERAEUS MEGAFUGE 16".
IT WAS REPORTED THAT BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES HAVE POOR SEPARATION BARRIER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PLASMA STILL BELOW GEL AFTER CENTRIFUGATION CENTRIFUGE SPEED 3500RPM FOR 1 MINUTES. CENTRIFUGE USED HERAEUS MEGAFUGE 16.