VERTEX RECONSTRUCTION SYSTEM
Report
- Report Number
- 1030489-2009-01006
- Event Type
- Injury
- Date Received
- October 29, 2009
- Report Date
- September 29, 2009
- Manufacturer
- WARSAW ORTHOPEDIC INC.
- Product Code
- KWP
- PMA / PMN Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE LOT OF THE SUSPECT DEVICE WAS NOT IDENTIFIED, THEREFORE, THE MANUFACTURER CANNOT DETERMINE THE SUSPECT DEVICE. FURTHER DETAILS ABOUT THE SCREWS ASSOCIATED WITH THIS EVENT. DEVICE HISTORY RECORDS FOR THESE LOTS WERE REVIEWED. NO DOCUMENTATION WAS FOUND THAT WOULD INDICATE A NON-CONFORMANCE TO SPECIFICATIONS. THIS PART IS NOT APPROVED FOR USE IN THE UNITED STATES; HOWEVER, A LIKE DEVICE CATALOG # 6955508, WAS CLEARED IN THE UNITED STATES. MANUFACTURE DATE FOR LOT W05K1083 IS 10/26/2005; MFR DATE FOR LOT W05K1086 IS 12/02/2005. NEITHER DEVICE NOR FILM OF APPLICABLE IMAGING STUDIES WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. WE ARE UNABLE TO DETERMINE THE CAUSE OF THE EVENT.
IT WAS REPORTED THAT THE PATIENT WHO HAD BASILAR IMPRESSION ASSOCIATED WITH RHEUMATISM UNDERWENT A PROCEDURE AT O-C3 USING POSTERIOR FIXATION. IT WAS REPORTED THAT ALL THREE CORTICAL SCREWS FOR THE MIDLINE OC PLATE BACKED OUT. THE PATIENT REPORTEDLY DEVELOPED A MODEST NUMBNESS. A REVISION SURGERY WAS PERFORMED APPROXIMATELY FIVE WEEKS POST OP. DURING THE REMOVING PROCEDURE, IT WAS FOUND THAT THE MIDLINE OC PLATE WAS IMPLANTED NOT AT MIDLINE BUT TOWARD RIGHT. THE PLATE WAS REPLACED. THE SURGEON BELIEVED THAT THERE WAS NOT ANY PROBLEM RELATED TO THE IMPLANTED PRODUCTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VERTEX RECONSTRUCTION SYSTEM | BONE SCREW | KWP | WARSAW ORTHOPEDIC INC. | NA | W05K1083 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Required Intervention | BONE SCREW CATALOG #: G6955508| LOT # W05K1083| LOT # W05K1086 |