FDA Adverse Event Malfunction Summary report: N

CAPTURE-R READY-SCREEN 3

MDR report key: 1518427 · Received October 30, 2009

Report

Report Number
1034569-2009-00379
Event Type
Malfunction
Date Received
October 30, 2009
Date of Event
September 30, 2009
Report Date
October 29, 2009
Manufacturer
IMMUCOR
Product Code
KSZ
PMA / PMN Number
STN 102707
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

REVIEW OF RESULT FILES- CRRS 3 LOT R056 WELL 1 (8) WELL 2 (40) WELL 3 (0) WELL 4 (100)- CONTROLS, VISUALLY OPTIMAL, WELL 1 SLIGHT HALO, SMALL BUTTON, WELL 2, MODERATE ADHERENCE WELL 3 -NEGATIVE PHENOTYPE- WELL 1JKA+JKB- WELL 2 JKA+JKB- WELL 3 JKA-JKB+CRRID LOT ID119RAW SCORES 1,2,1,2,2,8,1,1,0,1,1,40,1,1,100,0 - VISUAL INSPECTION CONTROLS, OPTIMAL. AGREE WITH INSTRUMENT ASSESSMENT. PHENOTYPE- WELL 2 JKA+JKB- WELL 4 JKA+JKB+ WELL 5 JKA+JKB+ WELL 6 JKA+JKB- WELL 11JKA+JKB+ WELL 12 JKA+JKB- WELL 13 JKA+JKB+ WELL 14 JKA+JKB+. THE REACTIVITY OF THE JKA ANTIGEN WAS CONFIRMED ON RETENTION CRRS(3), LOT R056 AND RETENTION CRRID, LOT ID119 WITH ANTI-JKA.THE REACTIVITY OF THE JKA ANTIGEN WAS CONFIRMED ON RETURNED CRRS(3), LOT R056 AND RETURNED CRRID, LOT ID119 WITH ANTI-JKA.THE SUBMITTED PLASMA SAMPLE (REPORTEDLY ANTI-JKA) WAS TESTED WITH RETURNED AND RETENTION CRRS(3), LOT R056 ON IN-HOUSE ECHO. THE SAMPLE EXHIBITED EQUIVOCAL, BUT VISUALLY POSITIVE REACTIVITY WITH CELL I AND CELL II, JK(A+) RED CELLS WITH RETURNED R056. THE SAMPLE EXHIBITED EQUIVOCAL, BUT VISUALLY POSITIVE REACTIVITY WITH CELL I AND 2+ REACTIVITY WITH CELL II, JK(A+) RED CELLS WITH RETENTION R056. A SERVICE CALL WAS MADE. THE ISSUE WAS RESOLVED BY REPLACING THE CAMERA READER. THE SYSTEM WAS TESTED AND OPERATES WITHIN SPECIFICATIONS WITH NO PROBLEMS OBSERVED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT A SAMPLE TESTED WITH CAPTURE-R READY SCREEN 3 (CRRS 3) AND CAPTURE-R READY ID (CRRID) WAS INTERPRETED AS NEGATIVE BY THE ECHO BUT WELLS APPEAR VISUALLY POSITIVE. CUSTOMER STATES THAT THE PATTERN FITS WITH A ANTI-JKA. CUSTOMER DOES NOT HAVE HISTORY ON THIS PATIENT. THIS IS THE FIRST TIME AN ANITBODY HAS BEEN IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPTURE-R READY-SCREEN 3 REAGENT RED BLOOD CELLS KSZ IMMUCOR R056

Patients

Seq Age Sex Outcome Treatment
1 69 YR