BD¿ BLUNT 5 MICRON FILTER NEEDLE 18 G X 1 1/2 IN
Report
- Report Number
- 1911916-2022-00408
- Event Type
- Malfunction
- Date Received
- August 8, 2022
- Date of Event
- July 12, 2022
- Report Date
- August 3, 2022
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- GAA
- UDI-DI
- 00382903052110
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: IT WAS REPORTED THERE WAS SOMETHING GREEN IN THE SYRINGE THAT LOOKS LIKE A PIECE OF PLASTIC. TO AID IN THE INVESTIGATION, TWO PHOTOS WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. ONE PHOTO SHOWS A BAYER SHELF BOX. THE OTHER PHOTO SHOWS A SYRINGE WITH A NEEDLE ASSEMBLY, INCLUDING THE PLASTIC SHIELD, CONNECTED TO IT. THE PRODUCT SHOWN IN THE PHOTO IS NOT A FILTER NEEDLE. AT THE LUER LOCK AREA OF THE SYRINGE BARREL THERE IS A RED CIRCLE INDICATING WHERE THE FOREIGN MATTER IS. THE PHOTO HAS A NOTE STATING THERE IS A GREEN SPECK. NO OTHER INFORMATION CAN BE OBTAINED FROM THE PHOTOS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 305211, LOT NUMBER 7200624. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED, BUT WITHOUT THE PHYSICAL SAMPLE ANALYSIS A PROBABLE ROOT CAUSE COULD NOT BE OFFERED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.
IT WAS REPORTED THAT THE BD¿ BLUNT 5 MICRON FILTER NEEDLE 18 G X 1 1/2 IN HAD FOREIGN MATTER IN DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BEFORE INJECTING THE PATIENT HE NOTICED THERE WAS SOMETHING GREEN IN THE SYRINGE THAT LOOKS LIKE A BIT OF PLASTIC. HE DID NOT INJECT THE PATIENT AND SENT THE SYRINGE BACK TO THE PHARMACY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2245416 | BD¿ BLUNT 5 MICRON FILTER NEEDLE 18 G X 1 1/2 IN | MANUAL SURGICAL INSTRUMENT FOR GENERAL USE | GAA | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 7200624 | 00382903052110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |