FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO

MDR report key: 1518200 · Received October 28, 2009

Report

Report Number
2953200-2009-01495
Event Type
Injury
Date Received
October 28, 2009
Date of Event
October 2, 2009
Report Date
October 2, 2009
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: RESULTS: ARTERIAL AND VENOUS OCCLUSION. DID NOT SLOWLY DEPLOY THE GRAFT BY ROTATING THE HANDLE. CONCLUSION: DID NOT SLOWLY DEPLOY THE GRAFT BY ROTATING THE HANDLE.

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED APPROX 31 MONTHS AGO, IN A PT AND PLACED INSIDE A DACRON TUBE GRAFT THAT WAS PREVIOUSLY IMPLANTED FOR AN OPEN REPAIR OF AN ILIAC AORTIC ANEURYSM. ANEURYSM AND VESSEL MORPHOLOGY WERE NOT REPORTED. IT WAS REPORTED THAT A CT SCAN OBTAINED APPROX 1 MONTH AGO, SHOWED THAT ANEURX STENT GRAFT HAD MIGRATED 2 CM WITH THE PRESENCE OF A PROXIMAL TYPE I ENDOLEAK (MFR # 2953200-2009-00229-1). THE AORTIC NECK HAD DILATED TO 28 MM OR 29 MM WITH AN ANGULATION OF 35 DEGREES. THERE WAS ILIAC FIXATION TO THE HYPOGASTRIC ARTERIES BILATERALLY. THE PHYSICIAN ELECTED TO TREAT THE MIGRATION WITH AN ANEURX CUFF AND A TALENT CUFF. IT WAS REPORTED THAT WHEN DEPLOYING THE ANEURX CUFF, THE PHYSICIAN PULLED THE TRIGGER ON THE DELIVERY SYSTEM AS SOON AS THE STENT GRAFT WAS POSITIONED, RATHER THAN SLOWLY DEPLOYING THE STENT GRAFT BY ROTATING THE HANDLE. CONSEQUENTLY, THE ANEURX CUFF ENDED UP COVERING BOTH RENAL ARTERIES. ATTEMPTS WERE MADE TO PULL THE CUFF DOWNWARD WITH BALLOONS AND WIRES THAT WERE ALREADY IN PLACE. THE CUFF MOVED SLIGHTLY DOWNWARDS ON THE RIGHT SIDE AND TILTED OVER, ON THE LEFT SIDE, THE CUFF WOULD NOT MOVE, BECAUSE IT WAS CATCHING ON THE BIFURCATED STENT GRAFT DISTALLY. THE PHYSICIAN ELECTED TO OPEN THE PT AND EXPLANT THE ANEURX CUFF; A DACRON GRAFT WAS THEN SEWN TO THE AORTA PROXIMALLY AND TO THE REMAINDER OF THE STENT GRAFT DISTALLY. THE ORIGINAL STENT GRAFTS WERE NOT EXPLANTED, AND THE EXPLANTED ANEURX CUFF WAS DISCARDED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00365182

Patients

Seq Age Sex Outcome Treatment
1 76 YR Required Intervention