FDA Adverse Event
Malfunction
Summary report: N
OMNIPOD INSULIN PUMP
MDR report key: 15180621
·
Received August 6, 2022
Report
- Report Number
- 3004464228-2022-13820
- Event Type
- Malfunction
- Date Received
- August 6, 2022
- Date of Event
- July 23, 2022
- Report Date
- July 30, 2022
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K192659
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR DETERMINE ITS ROOT CAUSE. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS.
Additional Manufacturer Narrative · 0
CORRECTION TO D(4): CATALOG NO CHANGED FROM UNAVAILABLE TO ZXP425. CORRECTION TO G(5): PMA/510(K) # CHANGED FROM UNAVAILABLE TO K192659.
Description of Event or Problem · 0
IT WAS REPORTED THE NEEDLE DEPLOYED EARLY, INDICATING A FAILURE OF THE NEEDLE MECHANISM. THE POD WAS NOT WORN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2275045 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 40160 | L50293 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR | Male |