FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER

MDR report key: 15178866 · Received August 5, 2022

Report

Report Number
1024879-2022-00428
Event Type
Malfunction
Date Received
August 5, 2022
Date of Event
June 29, 2022
Report Date
August 30, 2022
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686501
PMA / PMN Number
K982541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY MATERIAL #: 368650, LOT/BATCH #: 2061622. BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR DEFECTIVE LOCKING MECHANISM WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER SAFETY SHIELD BROKE OFF. NO USER OR PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE NEEDLE SAFETY COVER BROKE AWAY FROM THE HOUSING - COULD NOT BE SAFELY ENGAGED AFTER BLOOD COLLECTION."

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER SAFETY SHIELD BROKE OFF. NO USER OR PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE NEEDLE SAFETY COVER BROKE AWAY FROM THE HOUSING - COULD NOT BE SAFELY ENGAGED AFTER BLOOD COLLECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2434090 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON, DICKINSON & CO., (BD) 368650 2961622 50382903686501

Patients

Seq Age Sex Outcome Treatment
1 Unknown