FDA Adverse Event Malfunction Summary report: N

FIBEROPTIC LIGHT CABLE

MDR report key: 1517765 · Received October 16, 2009

Report

Report Number
1418479-2009-00013
Event Type
Malfunction
Date Received
October 16, 2009
Date of Event
September 17, 2009
Report Date
October 16, 2009
Manufacturer
RICHARD WOLF MEDICAL INSTRUMENTS
Product Code
EQH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY - 8061.456 - FIBEROPTIC LIGHT CABLE, 4.5MM - LOT#: 184O02. THE FIBEROPTIC LIGHT CABLE WAS CHECKED FOR LIGHT OUTPUT AND RESULTS WERE LOW LIGHT OUTPUT. THE ADAPTER ON THE CABLE WAS LOOSE AND THE RING WAS DISCOLORED. THERE WERE VISIBLE BROKEN FIBERS WITH BROWNISH COLOR. THE APPEARANCE OF THE CABLE SHOWS WEAR AND TEAR. THE CABLE HAS BEEN IN USE FOR APPROXIMATELY (7) SEVEN YEARS. THE CABLE WAS RETURNED TO THE CUSTOMER, PER THEIR REQUEST, AS IS. CAUSE: UNUSUAL, HOWEVER, WE BELIEVE THE BROKEN END FIBERS ALLOWED PROJECTED LIGHT TO SHINE THROUGH THE RUBBER INSULATION AND HEAT UP THE RUBBER MATERIAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE FIBER LIGHT CABLE CAUSED A 1.000 INCH X .250 INCH BURN ON THE PATIENT'S ABDOMEN. THE APPEARANCE WAS OF A 'BUBBLE' WITHOUT REDNESS. ANTIBIOTIC OINTMENT WAS APPLIED TO THE BURN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FIBEROPTIC LIGHT CABLE FIBEROPTIC LIGHT CABLE EQH RICHARD WOLF MEDICAL INSTRUMENTS 8061.456 184O02

Patients

Seq Age Sex Outcome Treatment
1 49 YR