CARB-BITE CRILE-WOOD NH 8
Report
- Report Number
- 3014334038-2022-00175
- Event Type
- Malfunction
- Date Received
- August 5, 2022
- Date of Event
- June 30, 2022
- Report Date
- September 7, 2022
- Manufacturer
- INTEGRA LIFESCIENCES MANSFIELD
- Product Code
- DWS
- UDI-DI
- 10381780377139
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- RISK MANAGER
- Health Professional
- N
Narratives
AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
THE CARB-BITE CRILE-WOOD NEEDLE HOODLER (121132) WAS NOT RETURNED FOR EVALUATION AFTER THREE GOOD FAITH ATTEMPTS (GFES) WERE MADE. LOT NUMBER INFORMATION HAS BEEN PROVIDED; THUS, DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES RELATED TO THE REPORTED ISSUE WERE IDENTIFIED. A DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. THE ISSUE OF BREAKING MAY BE THE RESULT OF ROUGH HANDLING OR ENVIRONMENTAL DAMAGE. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE IN THE FUTURE, THIS COMPLAINT WILL BE REOPENED, AND THE RESPECTIVE EVALUATION PERFORMED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR ISSUES. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.
THE FOLLOWING INFORMATION WAS REPORTED VIA UF/IMPORTER REPORT# (B)(4), RECEIVED ON 14-JULY-2022: WHILE SURGEON WAS WORKING, NEEDLE DRIVER BROKE IN THE WOUND. ALL PIECES FOUND AND VISUALIZED. X-RAY CALLED AND 2 VIEWS TAKEN AND THERE WAS NO FOREIGN BODY IN WOUND. ORIGINAL INTENDED PROCEDURE: OPEN REDUCTION INTERNAL FIXATION RIGHT DISTAL RADIUS AND SCAPHOID FRACTURE (RIGHT) NO PATIENT INJURY, DEATH OR SURGICAL DELAY WAS REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
N/A
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2275761 | CARB-BITE CRILE-WOOD NH 8 | CARDIOVASCULAR NEEDLE HOLDERS | DWS | INTEGRA LIFESCIENCES MANSFIELD | 121132 | 100084-1703 | 10381780377139 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male |