GORE® EXCLUDER® AAA ENDOPROSTHESIS
Report
- Report Number
- 2017233-2022-03189
- Event Type
- Injury
- Date Received
- August 4, 2022
- Date of Event
- July 14, 2022
- Report Date
- December 23, 2022
- Manufacturer
- W. L. GORE & ASSOCIATES, INC.
- Product Code
- MIH
- PMA / PMN Number
- P020004
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). W. L. GORE & ASSOCIATES, INC. (GORE) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY GORE, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. BLANK FIELDS PRESENT ON THIS REPORT INCLUDE REQUIRED FIELDS AND FIELDS DETERMINED TO BE NOT APPLICABLE. BLANK REQUIRED FIELDS INDICATE THAT THE INFORMATION WAS NOT PROVIDED, WAS DEEMED UNAVAILABLE OR WAS NOT APPLICABLE. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, GORE, OR ITS ASSOCIATES THAT THE DEVICE, GORE OR ITS ASSOCIATES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE A LEGAL ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED, AS DEFINED FROM A LEGAL STANDPOINT. THESE WORDS ARE INCLUDED IN THE REPORT AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
G CONTACT OFFICE - MANUFACTURING SITE: CORRECTED TO MEDICAL PHOENIX 2, 32470 N. NORTH VALLEY PARKWAY, PHOENIX AZ 85085,
ON (B)(6) 2015, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESIS. ON AN UNKNOWN DATE, DURING A FOLLOW UP EXAMINATION, ANEURYSM ENLARGEMENT WAS OBSERVED. ON (B)(6) 2022, REINTERVENTION WAS PERFORMED. A TYPE II ENDOLEAK WAS OBSERVED ON THE INTRAOPERATIVE ANGIOGRAPHY, LUMBER ARTERY AND RIGHT INTERNAL ILIAC ARTERY WERE EMBOLIZED. ALSO MIGRATIONS AT THE PROXIMAL SIDE OF THE TRUNK-IPSILATERAL LEG COMPONENT AND AT THE DISTAL SIDE OF THE CONTRALATERAL LEG COMPONENT WERE SUSPECTED, SO ADDITIONAL STENT GRAFTS WERE PLACED AT EACH SITE. THE PATIENT TOLERATED THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2602889 | GORE® EXCLUDER® AAA ENDOPROSTHESIS | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | W. L. GORE & ASSOCIATES, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Male | Required Intervention |