VASOVIEW ENDOSCOPIC VESSEL HARVESTING SYSTEM ACCESSORY PACK
Report
- Report Number
- 3006262933-2009-01309
- Event Type
- Malfunction
- Date Received
- October 22, 2009
- Date of Event
- September 23, 2009
- Report Date
- October 9, 2009
- Manufacturer
- MAQUET CARDIOVASCULAR
- Product Code
- GEI
- PMA / PMN Number
- K041981
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION RESULTS: BASED ON THE REPORTED COMPLAINT AND THE VISUAL ANALYSIS, THIS IS A CASE OF THE VALVE BACKING OUT OF THE LUER FITTING CAUSING THE BALLOON TO DEFLATE DUE TO A DEFECTIVE VALVE SUBASSEMBLY. THE REPORTED FAILURE "SHORT PORT BTT BLACK INFLATION VALVE DISLODGED" WAS CONFIRMED. (B) (4).
THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE SHORT PORT BTT BLACK INFLATION VALVE DISLODGED (POPPED OUT) WHEN STAFF PUSHED IN THE SYRINGE, FILLED THE BALLOON WITH AIR THEN PULLED OUT THE SYRINGE. THE NEXT THREE ACCESSORIES SHORT PORTS BTT WERE USED WITH THE SAME RESULTS. THE STAFF CONTINUED TO OPEN ACCESSORY KIT UNTIL THEY GOT A GOOD ONE TO USE AND THE PROCEDURE WAS COMPLETED. THE HOSP DID NOT REPORT ANY PT COMPLICATIONS. THIS REPORT IS FOR THE THIRD DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW ENDOSCOPIC VESSEL HARVESTING SYSTEM ACCESSORY PACK | GEI | MAQUET CARDIOVASCULAR | VH-2004 | 9072871 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |