BD 5ML SYRINGE LUER-LOK ¿ TIP
Report
- Report Number
- 9610847-2022-00281
- Event Type
- Malfunction
- Date Received
- August 4, 2022
- Date of Event
- July 5, 2022
- Report Date
- August 24, 2022
- Manufacturer
- BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
- Product Code
- FMF
- UDI-DI
- 00382903097036
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY OUR QUALITY ENGINEER INSPECTED THE 10 SAMPLES AND ONE PHOTOGRAPH SUBMITTED FOR EVALUATION. THE REPORTED ISSUE OF PLUNGER ROD BROKEN / DAMAGED WAS CONFIRMED UPON INSPECTION OF THE SAMPLES. FIVE LOOSE SAMPLES FROM BATCH 2090479 HAVE BROKEN PLUNGER RODS ON FOUR OF THE SYRINGES AND ONE OF THE SYRINGES HAS CRACKED GATE BLOCK WITHIN ACCEPTABLE LIMITS FROM THE ROOF OF THE BARREL UP TO THE BARREL TIP. THE IMAGE SHOWS 2 PLUNGER RODS INSIDE THE BARRELS WITH BOTH PLUNGER RODS CRACKED IN TWO PLACES FROM AN UNKNOWN BATCH #. BD DETERMINED THAT THE POTENTIAL ROOT CAUSE FOR THE CRACKED PLUNGER ROD DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. SEE H10.
THIS MDR IS THE FIRST OF 3 RELATED MDRS AND ADDRESSES THE LOT # 2090479 EVENT. IT WAS REPORTED THAT 19 BD 5ML SYRINGE LUER-LOK ¿ TIP HAD CRACKED PLUNGER RODS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CUSTOMER REPORTED SYRINGES WITH CRACKED PLUNGER RODS.
THIS MDR IS THE FIRST OF 3 RELATED MDRS AND ADDRESSES THE LOT # 2090479 EVENT. IT WAS REPORTED THAT 19 BD 5ML SYRINGE LUER-LOK ¿ TIP HAD CRACKED PLUNGER RODS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CUSTOMER REPORTED SYRINGES WITH CRACKED PLUNGER RODS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2111407 | BD 5ML SYRINGE LUER-LOK ¿ TIP | PISTON SYRINGE | FMF | BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. | 2090479 | 00382903097036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |