CHECK-IT
Report
- Report Number
- 3016521623-2022-00266
- Event Type
- Malfunction
- Date Received
- August 4, 2022
- Date of Event
- July 21, 2022
- Report Date
- August 1, 2022
- Manufacturer
- LUCIRA HEALTH, INC.
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT HAS NOT BEEN RETURNED. A DHR REVIEW OF THE ASSOCIATED KIT LOT NUMBER (K08A112203223M1) WAS COMPLETED, AND 0 DISCREPANCIES WERE FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THIS LOT. THERE IS ONE SIMILAR COMPLAINT (B)(4) WITH AN ALLEGED FALSE POSITIVE REPORT FROM A DIFFERENT CUSTOMER ASSOCIATED WITH THE SAME KIT LOT NUMBER K08A112203223M1. AN INVESTIGATION INTO THESE COMPLAINTS HAVE BEEN REVIEWED TO OBTAIN ADDITIONAL INFORMATION. BASED ON REVIEW OF THE PRODUCT'S FMEAS AND RISK ASSESSMENT DOCUMENTATION, FALSE POSITIVE TEST RESULTS ARE A KNOWN POSSIBLE OUTCOME REGARDING THIS ISSUE UNDER EVALUATION. THE COMPLAINT RATE FOR "FALSE POSITIVE" IS UNDER THE EXPECTED THRESHOLD OF 2% (LABEL CLAIM)/1% INTERNAL WARNING LIMIT. LUCIRA HEALTH WILL CONTINUE TO MONITOR TRENDS RELATED TO FALSE POSITIVE RESULTS IN ACCORDANCE WITH POST-MARKET SURVEILLANCE PROCESS. DHR REVIEW KIT LOT# K08A112203223M1: SAMPLE VIAL LOT DHR REVIEW: 2203138 (ASSOCIATED INTERNAL LOT # 220124-4V) TEST LOT DHRS REVIEW: 2202448, 2202476, 2202475, 2202473, 2202476, 2202475, 2202474, 2202394, 2203073, 2203081, 2203072 (ASSOCIATED INTERNAL LOT # 220124-27U, 220124-30L, 220124-30I, 220124-30C, 220124-30K, 220124-30G, 220124-30F, 220124-12N, 220305-3E, 220305-3G, 220305-3C) A FAILURE ANALYSIS (FA) WAS CONDUCTED FOR KIT LOT # K08A112203223M1 ON MAY 18, 2022, ASSOCIATED WITH COMPLAINT CP2963: 3016521623-2022-00165 THROUGH 3016521623-2022-00176 FOR FURTHER INVESTIGATION (ALL ASSOCIATED KIT LOT NUMBERS UNDERWENT FURTHER INVESTIGATION TESTING DUE TO AN ALLEGED FALSE POSITIVE REPORT). FAILURE ANALYSIS TESTING FOR KIT LOT# K08A112203223M1 RESULTS: LUCIRA HEALTH PULLED RETAINS ASSOCIATED WITH KIT LOT NUMBER K08A112203223M1 (50 TEST UNITS, 50 SAMPLE VIALS) FOR TESTING. THE TEST UNITS/SAMPLE VIALS WERE TESTED FOLLOWING WI162 (DEVICE LOT RELEASED-NO TEMPLATE CONTROL (NTC) TESTING) SECTION 9.3. ALL RETAINED DEVICES PASSED WITHOUT ANY FALSE POSITIVE RESULTS. THEREFORE, KIT LOT NUMBER K08A112203223M1 PASSED THE RELEASE CRITERIA IN THE INTERNAL TESTING. BASED ON THE LIMITED INFORMATION AVAILABLE, ROOT CAUSE CANNOT BE DETERMINED. POTENTIAL ROOT CAUSES INCLUDE BUT ARE NOT LIMITED TO: ENVIRONMENTAL CONTAMINATION, LOW VIRAL LOAD, AND/OR DEVICE FAILURE. A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER (B)(4) CHECK-IT.
ONE DEVICE REPORTED AS HAVING AN ALLEGED FALSE POSITIVE RESULT. THE CUSTOMER PERFORMED MULTIPLE LUCIRA TESTS WITH ONE RESULTING IN A POSITIVE. THE CUSTOMER FOLLOWED-UP WITH PERFORMING TWO ADDITIONAL LUCIRA TEST KITS (ALL FROM ONE LOT BATCH) WITH NEGATIVE RESULTS FOR CONFIRMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2493486 | CHECK-IT | COVID-19 TEST KIT | QJR | LUCIRA HEALTH, INC. | K08A112203223M1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |