FDA Adverse Event Injury Summary report: N

BLUNT TIP SCREW, 4X42MM

MDR report key: 15161935 · Received August 4, 2022

Report

Report Number
0009613350-2022-00403
Event Type
Injury
Date Received
August 4, 2022
Date of Event
July 7, 2022
Report Date
September 9, 2022
Manufacturer
ZIMMER SWITZERLAND MANUFACTURING GMBH
Product Code
HSB
UDI-DI
00889024505452
PMA / PMN Number
K181827
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CONCOMITANT MEDICAL PRODUCTS: PROXIMAL HUMERUS, LEFT, 11X160MM, CATALOG:47-2496-161-11, LOT: 3068440. BLUNT TIP SCREW, 4X44MM, CATALOG: 47-2486-044-40, LOT: 3024704. BLUNT TIP SCREW, 4X44MM, CATALOG: 47-2486-044-40, LOT: 3054485. BLUNT TIP SCREW, 4X52MM, CATALOG: 47-2486-052-40, LOT: UNKNOWN. ANN CORT BONE SCREW 4 X 28MM, CATALOG: 47-2486-128-40, LOT: 3081901. ANN CORT BONE SCREW 4 X 30MM, 4X28MM, CATALOG: 47-2486-130-40, LOT: 3073753. AFFIXUS PH NL CAP 0MM, CATALOG: 47-2488-010-00, LOT: 3081992. TORQUE LIMITING HANDLE,: CATALOG: 27923, LOT: UNKNOWN. REPORT SOURCE - FOREIGN: JAPAN. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0009613350-2022-00401, 0009613350-2022-00409, 0009613350-2022-00410, 0009613350-2022-00411.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. AS NO PRODUCT WAS RETURNED, VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. DEVICES ARE USED FOR TREATMENT. NO SURGICAL REPORT IS AVAILABLE FOR INVESTIGATION. THERE IS ONE UNDATED X-RAY PICTURE AVAILABLE. THE PICTURE DEMONSTRATES BACKING OUT OF THE PROXIMAL FIXATION SCREWS. BASED ON THE INVESTIGATION IT COULD BE ASSUMED THAT FURTHER POSSIBLE CONTRIBUTING FACTORS TO THE MIGRATION OF THE SCREW MIGHT BE MULTIFACTORIAL RELATED TO EITHER PATIENT CONDITION, BEHAVIOUR OR IMPLANTATION PROCEDURE. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. THE INVESTIGATION ALSO CONSIDERED WHETHER THE DESIGN LIMITATION OF THE CORELOCK MECHANISM COULD BE A POTENTIAL CONTRIBUTING FACTOR. HOWEVER, FURTHER BIOMECHANICAL TESTING IDENTIFIED THAT THE PRODUCT PERFORMANCE IS IN AN ACCEPTABLE RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED WITH ANN NAIL. AFTER 2 MONTHS FROM THE INITIAL SURGERY, THE SURGEON FOUND 2 PROXIMAL SCREW WAS BACKED OUT FROM THE PROPER POSITION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2063955 BLUNT TIP SCREW, 4X42MM ROD, FIXATION HSB ZIMMER SWITZERLAND MANUFACTURING GMBH N/A 3076740 00889024505452

Patients

Seq Age Sex Outcome Treatment
1 Unknown Hospitalization SEE H10 NARRATIVE.