PROXIMAL HUMERUS, LEFT, 11X160MM
Report
- Report Number
- 0009613350-2022-00401
- Event Type
- Injury
- Date Received
- August 4, 2022
- Date of Event
- July 7, 2022
- Report Date
- September 9, 2022
- Manufacturer
- ZIMMER SWITZERLAND MANUFACTURING GMBH
- Product Code
- HSB
- UDI-DI
- 00889024505827
- PMA / PMN Number
- K181827
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). CONCOMITANT MEDICAL PRODUCTS: BLUNT TIP SCREW, 4X42MM, CATALOG:47-2486-042-40, LOT: 3076740. BLUNT TIP SCREW, 4X44MM, CATALOG: 47-2486-044-40, LOT: 3024704. BLUNT TIP SCREW, 4X44MM, CATALOG: 47-2486-044-40, LOT: 3054485. BLUNT TIP SCREW, 4X52MM, CATALOG: 47-2486-052-40, LOT: UNKNOWN. ANN CORT BONE SCREW 4 X 28MM, CATALOG: 47-2486-128-40, LOT: 3081901. ANN CORT BONE SCREW 4 X 30MM, 4X28MM, CATALOG: 47-2486-130-40, LOT: 3073753. AFFIXUS PH NL CAP 0MM, CATALOG: 47-2488-010-00, LOT: 3081992. TORQUE LIMITING HANDLE, CATALOG:27923, LOT: UNKNOWN. REPORT SOURCE - FOREIGN: JAPAN. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0009613350-2022-00403, 0009613350-2022-00409, 0009613350-2022-00410, 0009613350-2022-00411.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. AS NO PRODUCT WAS RETURNED, VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. DEVICES ARE USED FOR TREATMENT. NO SURGICAL REPORT IS AVAILABLE FOR INVESTIGATION. THERE IS ONE UNDATED X-RAY PICTURE AVAILABLE. THE PICTURE DEMONSTRATES BACKING OUT OF THE PROXIMAL FIXATION SCREWS. BASED ON THE INVESTIGATION IT COULD BE ASSUMED THAT FURTHER POSSIBLE CONTRIBUTING FACTORS TO THE MIGRATION OF THE SCREW MIGHT BE MULTIFACTORIAL RELATED TO EITHER PATIENT CONDITION, BEHAVIOUR OR IMPLANTATION PROCEDURE. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. THE INVESTIGATION ALSO CONSIDERED WHETHER THE DESIGN LIMITATION OF THE CORELOCK MECHANISM COULD BE A POTENTIAL CONTRIBUTING FACTOR. HOWEVER, FURTHER BIOMECHANICAL TESTING IDENTIFIED THAT THE PRODUCT PERFORMANCE IS IN AN ACCEPTABLE RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED WITH ANN NAIL. AFTER 2 MONTHS FROM THE INITIAL SURGERY, THE SURGEON FOUND 2 PROXIMAL SCREW WAS BACKED OUT FROM THE PROPER POSITION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1704930 | PROXIMAL HUMERUS, LEFT, 11X160MM | ROD, FIXATION | HSB | ZIMMER SWITZERLAND MANUFACTURING GMBH | N/A | 3068440 | 00889024505827 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Hospitalization |