BBL - MUELLER HINTON AGAR WITH 5% SHEEP BLOOD
Report
- Report Number
- 1119779-2022-01056
- Event Type
- Malfunction
- Date Received
- August 3, 2022
- Date of Event
- July 7, 2022
- Report Date
- August 31, 2022
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- JTZ
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. THE MANUFACTURING LOCATION FOR THIS PRODUCT IS FUKUSHIMA, JAPAN. THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, SPARKS HAS BEEN LISTED AND THE SPARKS FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER.
H.6. INVESTIGATION SUMMARY 251801 LOT#2073206 CONTAMINATION WE CONFIRMED THE ISSUE AS A REPORT FROM RETURNED SAMPLE. THE ISSUE WAS CONTAMINATION. NO TREND. WE WILL CONTINUE TO MONITOR THIS ISSUE.
IT WAS REPORTED THAT WHILE USING BBL - MUELLER HINTON AGAR WITH 5% SHEEP BLOOD THE CUSTOMER NOTICED BIOLOGICAL CONTAMINATION . THIS EVENT OCCURRED 1 TIME. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE CUSTOMER FOUND FOREIGN MATTER WHICH COULD BE BACTERIA IN THE MEDIA."
IT WAS REPORTED THAT WHILE USING BBL - MUELLER HINTON AGAR WITH 5% SHEEP BLOOD THE CUSTOMER NOTICED BIOLOGICAL CONTAMINATION . THIS EVENT OCCURRED 1 TIME. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE CUSTOMER FOUND FOREIGN MATTER WHICH COULD BE BACTERIA IN THE MEDIA."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1903874 | BBL - MUELLER HINTON AGAR WITH 5% SHEEP BLOOD | CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH | JTZ | BECTON, DICKINSON & CO. (SPARKS) | 2073206 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |