CTR34, 11X100 KII OPT ZTHR HDL 6/BX
Report
- Report Number
- 2027111-2022-00683
- Event Type
- Malfunction
- Date Received
- August 2, 2022
- Report Date
- September 22, 2022
- Manufacturer
- APPLIED MEDICAL RESOURCES
- Product Code
- GCJ
- UDI-DI
- 00607915124243
- PMA / PMN Number
- K060096
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE EVENT DEVICE IS ANTICIPATED TO BE RETURNED TO APPLIED MEDICAL FOR EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED UPON COMPLETION OF THE INVESTIGATION.
THE EVENT UNIT WAS RETURNED TO APPLIED MEDICAL FOR EVALUATION. VISUAL AND FUNCTIONAL TESTING WAS PERFORMED ON THE EVENT UNIT, HOWEVER, THE COMPLAINANT¿S EXPERIENCE OF IMPROPER FIT COULD NOT BE REPLICATED OR CONFIRMED. THE EVENT UNIT MET CURRENT SPECIFICATIONS AND THERE WERE NO VISIBLE NON-CONFORMANCES. APPLIED MEDICAL HAS REVIEWED THE DETAILS SURROUNDING THE EVENT AND RELATED PRODUCTS AND IS UNABLE TO DETERMINE THE CAUSE OF THE REPORTED EVENT BASED ON THE EVALUATION OF THE RETURNED UNIT. THE PROBABILITY AND CRITICALITY OF HARM RESULTING FROM THIS FAILURE HAVE BEEN EVALUATED AND WERE FOUND TO BE AT AN ACCEPTABLE LEVEL. THIS EVENT IS NOT REPORTABLE AS IT IS UNLIKELY TO CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY. CORRECTION: UPDATED SECTION H6 COMPONENT CODES TO 887 - MOUNT.
PROCEDURE PERFORMED: LAPAROSCOPIC PROCEDURES. EVENT DESCRIPTION: EVENT 1 OF 5: 2022-001526 EVENT. 2 OF 5: 2022-001535. EVENT 3 OF 5: 2022-001536. EVENT 4 OF 5: 2022-001537. EVENT 5 OF 5: 2022-001538. DATE OF EVENT: (B)(6) 2022. THE SURGEON USE THE OPTICAL TECNIC FOR THE FIRST ACCESS IN LAPAROSCOPIC PROCEDURES. DURING 2 DIFFERENT PROCEDURES THE OPTURATOR OF THE TROCAR IS TOO SMALL TO HOUSE THE LAPAROSCOPE. THE SURGEON INFORM THAT THE 5 TROCARS HAD THE SAME PROBLEM, BUT SHE DOESN'T REMEMBER THE MODEL AND THE LOT OF THIS TROCARS. ADDITIONAL INFORMATION RECEIVED BY EMAIL FROM APPLIED MEDICAL REPRESENTATIVE ON (B)(6) 2022: MODEL CTR34 LOT 1424056. ADDITIONAL INFORMATION RECEIVED BY EMAIL FROM APPLIED MEDICAL REPRESENTATIVE ON (B)(6) 2022: BRAND OF LAPAROSCOPE: [BRAND NAME] MODEL OF THE LAPAROSCOPE: 30° OPTICS SIZE: 10MM. ADDITIONAL INFORMATION RECEIVED BY EMAIL FROM APPLIED MEDICAL REPRESENTATIVE ON (B)(6) 2022: THE USER EXPERIENCED INCONSISTENT RESISTANCE DURING INSTRUMENT MANIPULATION. PATIENT STATUS: NO PATIENT INJURY HAS BEEN REPORTED.
PROCEDURE PERFORMED: LAPAROSCOPIC PROCEDURES EVENT DESCRIPTION: EVENT 1 OF 5: (B)(4) (MFR REPORT 2027111-2022-00679); EVENT 2 OF 5: (B)(4) (MFR REPORT 2027111-2022-00680); EVENT 3 OF 5: (B)(4) (MFR REPORT 2027111-2022-00681); EVENT 4 OF 5: (B)(4) (MFR REPORT 2027111-2022-00682); EVENT 5 OF 5: (B)(4) (MFR REPORT 2027111-2022-00683). DATE OF EVENT: 30-31/05/2022. THE SURGEON USE THE OPTICAL TECNIC FOR THE FIRT ACCES IN LAPAROSCOPIC PROCEDURES . DURING 2 DIFFERENT PROCEDURES THE OPTURATOR OF THE TROCAR IS TOO SMALL TO HOUSE THE LAPAROSCOPE. THE SURGEON INFORM THAT THE 5 TROCARS HAD THE SAME PROBLEM, BUT SHE DON'T REMEMBER THE MODEL AND THE LOT OF THIS TROCARS. ADDITIONAL INFORMATION RECEIVED BY EMAIL FROM APPLIED MEDICAL REPRESENTATIVE ON 15JUNE22: MODEL CTR34 LOT 1424056. ADDITIONAL INFORMATION RECEIVED BY EMAIL FROM APPLIED MEDICAL REPRESENTATIVE ON 1JULY2022: BRAND OF LAPAROSCOPE: [BRAND NAME] MODEL OF THE LAPAROSCOPE: 30° OPTICS SIZE: 10MM. ADDITIONAL INFORMATION RECEIVED BY EMAIL FROM APPLIED MEDICAL REPRESENTATIVE ON 12JULY22: THE USER EXPERIENCED INCONSISTENT RESISTANCE DURING INSTRUMENT MANIPULATION. PATIENT STATUS: NO PATIENT INJURY HAS BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 952626 | CTR34, 11X100 KII OPT ZTHR HDL 6/BX | LAPAROSCOPE, GENERAL & PLASTIC SURGERY | GCJ | APPLIED MEDICAL RESOURCES | CTR34 | 1424056 | 00607915124243 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | [BRAND NAME] LAPAROSCOPE |