FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 15150352 · Received August 2, 2022

Report

Report Number
1000113657-2022-00425
Event Type
Malfunction
Date Received
August 2, 2022
Date of Event
July 8, 2022
Report Date
August 29, 2022
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
000212920091
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 29-AUG-2022: H6: UPDATED FDA¿S TYPE, FINDINGS AND CONCLUSIONS CODES. H10: METER WAS NOT RETURNED FOR EVALUATION. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 174, 173, 185, 162 AND 185 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 97-108 MG/DL AND EXPECTED PM NON-FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 140-151 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULT OF 176 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE LIVING ROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 09/30/2023 AND OPEN VIAL DATE IS ONE WEEK AGO. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1 : 174 MG/DL DATE: (B)(6) TIME: 03:23 AM FASTING, RESULT 2 : 173 MG/DL DATE: (B)(6) TIME: 03:31 AM FASTING, RESULT 3 : 185 MG/DL DATE: (B)(6) TIME: 03:30 AM FASTING, RESULT 4 : 162 MG/DL DATE: (B)(6) TIME: 06:39 PM NON-FASTING, RESULT 5 : 185 MG/DL DATE: (B)(6) TIME: 04:14 AM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2064237 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX HEB 30CT24/CASE MG/DL ZA4716S 000212920091

Patients

Seq Age Sex Outcome Treatment
1 Unknown