FDA Adverse Event Malfunction Summary report: N

HD AUTOCLAVABLE CAMERA HEAD

MDR report key: 15148227 · Received August 2, 2022

Report

Report Number
3002808148-2022-00569
Event Type
Malfunction
Date Received
August 2, 2022
Date of Event
July 3, 2022
Report Date
September 2, 2022
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
UDI-DI
04953170310799
PMA / PMN Number
K102059
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS EVALUATED BY OLYMPUS. THE EVALUATION FOUND NO IMAGE WAS PRESENT DUE TO A CABLE FAILURE. THE PROBLEM REPORTED BY THE USER FACILITY COULD NOT BE CONFIRMED. THE INVESTIGATION IS ONGOING AND THE DEFINITIVE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED ACCORDINGLY.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION AND CORRECTION TO H4. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE OF THE DEFECTIVE CABLE COULD NOT BE IDENTIFIED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

A USER FACILITY RETURNED THE OLYMPUS, MODEL CH-S190-XZ-E, HD AUTOCLAVABLE CAMERA HEAD TO OLYMPUS FOR REPAIR DUE TO A REPORT OF "POOR CONTACT OF THE CAMERA CONICAL SHEATH." THE PROBLEM, AS REPORTED TO OLYMPUS, WAS IDENTIFIED DURING REPROCESSING. THERE WAS NO PATIENT INJURY, ASSOCIATED WITH THE PROBLEM, REPORTED TO OLYMPUS. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, THE IMAGE DID NOT APPEAR. THIS REPORT IS SUBMITTED DUE TO THE FINDING OF NO IMAGE PRESENT, IDENTIFIED DURING IN-HOUSE SERVICE OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2080778 HD AUTOCLAVABLE CAMERA HEAD AUTOCLAVABLE CAMERA HEAD FET SHIRAKAWA OLYMPUS CO., LTD. CH-S190-XZ-E 04953170310799

Patients

Seq Age Sex Outcome Treatment
1 Unknown