FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 15138263 · Received August 1, 2022

Report

Report Number
1314492-2022-03227
Event Type
Malfunction
Date Received
August 1, 2022
Report Date
September 1, 2022
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
UDI-DI
00085412498683
PMA / PMN Number
K133801
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

H10: THE DEVICE WAS RECEIVED FOR EVALUATION. DURING FUNCTIONAL FLOW ACCURACY TESTING, AN UNDERINFUSION WITH RESULTS OUT OF THE ACCEPTED RANGE WAS NOT REPRODUCED. A SERVICE HISTORY REVIEW REVEALED NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE REPORTED CONDITION WAS NOT VERIFIED. THERE IS CURRENTLY AN OPEN FIELD ACTION (FA-2021-056) ASSOCIATED WITH REINFORCING PROPER INTRAVENOUS LINE SETUP AND DETECTION OF UPSTREAM OCCLUSION FOR SPECTRUM PUMPS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SPECTRUM PUMP UNDERINFUSED DURING TESTING IN THE BIOMEDICAL SERVICE DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2935806 SPECTRUM INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION NA NA 00085412498683

Patients

Seq Age Sex Outcome Treatment
1 Unknown