SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2022-03227
- Event Type
- Malfunction
- Date Received
- August 1, 2022
- Report Date
- September 1, 2022
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- UDI-DI
- 00085412498683
- PMA / PMN Number
- K133801
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
H10: THE DEVICE WAS RECEIVED FOR EVALUATION. DURING FUNCTIONAL FLOW ACCURACY TESTING, AN UNDERINFUSION WITH RESULTS OUT OF THE ACCEPTED RANGE WAS NOT REPRODUCED. A SERVICE HISTORY REVIEW REVEALED NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE REPORTED CONDITION WAS NOT VERIFIED. THERE IS CURRENTLY AN OPEN FIELD ACTION (FA-2021-056) ASSOCIATED WITH REINFORCING PROPER INTRAVENOUS LINE SETUP AND DETECTION OF UPSTREAM OCCLUSION FOR SPECTRUM PUMPS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SPECTRUM PUMP UNDERINFUSED DURING TESTING IN THE BIOMEDICAL SERVICE DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2935806 | SPECTRUM INFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE CORPORATION | NA | NA | 00085412498683 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |