FDA Adverse Event Malfunction Summary report: N

GENESIS II SPC F/E SZ GDE

MDR report key: 15136780 · Received July 30, 2022

Report

Report Number
1020279-2022-03563
Event Type
Malfunction
Date Received
July 30, 2022
Date of Event
July 18, 2022
Report Date
September 1, 2022
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HRY
UDI-DI
00885556059951
PMA / PMN Number
K121393
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE (B)(4).

Additional Manufacturer Narrative · 0

H3, H6: THE ASSOCIATED DEVICE WAS RETURNED AND EVALUATED. A VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMS THE CONNECTOR BOLT IS BROKEN, RENDERING THE DEVICE INOPERABLE. THE DEVICE SHOWS SIGNS OF SIGNIFICANT WEAR AND USE. A REVIEW OF COMPLAINT HISTORY REVEALED SIMILAR EVENTS FOR THE LISTED DEVICE OVER THE PREVIOUS 12 MONTHS, BUT NO SIMILAR EVENTS FOR THE BATCH BASED ON THE HISTORICAL DATA, THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. AT THIS TIME, WE DO NOT HAVE REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THE ANTICIPATED RISK LEVEL IS STILL ADEQUATE. ASSESSMENT OF HISTORICAL ESCALATED CASES CONCLUDED THAT THERE ARE NO PRIOR ACTIONS RELATED TO THIS DEVICE AND FAILURE MODE. AS THE DEVICE BROKE AND IT CAN NO LONGER FIT ITS PURPOSE, THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD BE CORROBORATED. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Additional Manufacturer Narrative · 0

LOT NUMBER UPDATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT, DURING A TKA SURGERY, THE GENESIS II SPC F/E SZ GDE BROKE. NO PIECES FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING A S+N BACK-UP DEVICE. NO FURTHER COMPLICATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1430554 GENESIS II SPC F/E SZ GDE PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER HRY SMITH & NEPHEW, INC. 71933043 14FCT0012 00885556059951

Patients

Seq Age Sex Outcome Treatment
1 Unknown