BD PLASTIPAK¿ 3-PIECE SYRINGE
Report
- Report Number
- 3003152976-2022-00340
- Event Type
- Malfunction
- Date Received
- July 29, 2022
- Date of Event
- June 28, 2022
- Report Date
- August 22, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVALUTION: YES. D10: RETURNED TO MANUFACTURER ON: 02-AUG-2022 H6: INVESTIGATION SUMMARY ONE SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUALLY INSPECTING THE PHOTO, A HAIR WAS OBSERVED INSIDE THE BLISTER PACK. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 2203036, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. WHILE WE COULD NOT IDENTIFY A DIRECT ISSUE, THE ROOT CAUSE OF THIS INCIDENT IS DUE TO PERSONNEL NOT FOLLOWING THE PROPER GOWNING PROCEDURE.
IT WAS REPORTED THAT A HAIR WAS FOUND IN THE BD PLASTIPAK¿ 3-PIECE SYRINGE UNIT PACKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "A DARK HAIR IN THE PACKAGE. THERE IS NO CHANCE THAT THE HAIR HAS GOTTEN THERE AFTER OPENING THE SYRINGE."
IT WAS REPORTED THAT A HAIR WAS FOUND IN THE BD PLASTIPAK¿ 3-PIECE SYRINGE UNIT PACKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "A DARK HAIR IN THE PACKAGE. THERE IS NO CHANCE THAT THE HAIR HAS GOTTEN THERE AFTER OPENING THE SYRINGE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 951455 | BD PLASTIPAK¿ 3-PIECE SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2203036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |