COBAS INTEGRA 800
Report
- Report Number
- 1823260-2009-07038
- Event Type
- Malfunction
- Date Received
- October 13, 2009
- Date of Event
- September 23, 2009
- Report Date
- October 13, 2009
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JGS
- PMA / PMN Number
- K951595
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 1: 127, 135. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 10: 131, 138. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 11 (MALE): 124, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 12: 125, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 13: 128, 139. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 14 (MALE): 131, 140. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 15: 133, 140. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 16: 123, 133. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 17: 126, 137. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 18: 131, 142. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 19: 112, 1110, 119. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 2: 122, 132. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 20 (FEMALE): 126, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 21: 124, 137. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 22: 125, 137. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 23: 121, 132. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 24: 127, 139. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 25: 128, 137. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 26: 126, 138. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 27: 124, 134. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 28: 122, 135. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 29: 124, 135. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 3: 132, 139. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 30: 132, 144. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 31: 127, 137. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 32: 125, 137. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 33: 122, 129. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 34: 126, 132. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 35: 123, 131. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 36: 125, 138. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 37: 123, 1316. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 38: 123, 134. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 39: 124, 139. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 4: 127, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 40: 129, 129, 141. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 41: 129, 144, 144. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 42: 123, 132. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 43 (FEMALE): 124, 125, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 44: 122, 122, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 45: 118, 134. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 46: 123, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 47: 117, 118, 129. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 48: 136, 125, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 49: 123, 137. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 5: 126, 136. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 50: 129, 143. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 51: 128, 140. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 52: 115, 125, 126. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 6: 128, 140. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 7: 131, 137. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 8: 130, 138. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
THE FIELD SERVICE REP DETERMINED THE DRY PRESSURE WAS LOW, CAUSING THE AIR FLOW VALVE TO BE ERRATIC. HE REPLACED AND ADJUSTED THE AIR FLOW NEEDLE VALVE. TO VERIFY THE ANALYZER PERFORMANCE, HE RAN CALIBRATION AND QC WITH ACCEPTABLE RESULTS.
THE USER RECEIVED ERRONEOUS ISE RESULTS FOR MULTIPLE PT SAMPLES AND STATED SHE HAD RECEIVED CALLS FROM THE NURSING STATIONS SUSPECTING ERRONEOUS SODIUM RESULTS. THE INITIAL RESULTS WERE RUN ON THE INTEGRA 800 (B)(4) ON (B)(6) 2009 AND WERE REPEATED ON OTHER INTEGRA 800 (B)(4) ON (B)(6) 2009. ALL RESULTS ARE IN MMOL PER L. OF THE PT DATA PROVIDED, 52 RESULTS WERE DISCREPANT. PT 9: 129, 139. THE ERRONEOUS RESULTS WERE REPORTED. NO PTS WERE ADVERSELY AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS INTEGRA 800 | CLINICAL CHEMISTRY ANALYZER - JJE | JGS | ROCHE DIAGNOSTICS | I800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | |||
| 10 | UNK | |||
| 11 | UNK | |||
| 12 | UNK | |||
| 13 | UNK | |||
| 14 | UNK | |||
| 15 | UNK | |||
| 16 | UNK | |||
| 17 | UNK | |||
| 18 | UNK | |||
| 19 | UNK | |||
| 2 | UNK | |||
| 20 | UNK | |||
| 21 | UNK | |||
| 22 | UNK | |||
| 23 | UNK | |||
| 24 | UNK | |||
| 25 | UNK | |||
| 26 | UNK | |||
| 27 | UNK | |||
| 28 | UNK | |||
| 29 | UNK | |||
| 3 | UNK | |||
| 30 | UNK | |||
| 31 | UNK | |||
| 32 | UNK | |||
| 33 | UNK | |||
| 34 | UNK | |||
| 35 | UNK | |||
| 36 | UNK | |||
| 37 | UNK | |||
| 38 | UNK | |||
| 39 | UNK | |||
| 4 | UNK | |||
| 40 | UNK | |||
| 41 | UNK | |||
| 42 | UNK | |||
| 43 | UNK | |||
| 44 | UNK | |||
| 45 | UNK | |||
| 46 | UNK | |||
| 47 | UNK | |||
| 48 | UNK | |||
| 49 | UNK | |||
| 5 | UNK | |||
| 50 | UNK | |||
| 51 | UNK | |||
| 52 | UNK | |||
| 6 | UNK | |||
| 7 | UNK | |||
| 8 | UNK | |||
| 9 | UNK |