T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2022-79328
- Event Type
- Malfunction
- Date Received
- July 29, 2022
- Date of Event
- July 8, 2022
- Report Date
- July 9, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00853052007264
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED AND EVALUATION IS PERFORMED. PER TANDEM USER GUIDE: AVOID SUBMERGING YOUR PUMP IN FLUID BEYOND A DEPTH OF 3 FEET (0.91 M) OR FOR MORE THAN 30 MINUTES (IPX7 RATING). IF YOUR PUMP HAS BEEN EXPOSED TO FLUID BEYOND THESE LIMITS, CHECK FOR ANY SIGNS OF FLUID ENTRY. IF THERE ARE SIGNS OF FLUID ENTRY, DISCONTINUE USE OF THE PUMP AND CONTACT CUSTOMER TECHNICAL SUPPORT. PER TANDEM USER GUIDE: DO NOT USE YOUR PUMP IF YOU THINK IT MIGHT BE DAMAGED DUE TO DROPPING IT OR HITTING IT AGAINST A HARD SURFACE. CHECK THAT THE PUMP IS WORKING PROPERLY BY PLUGGING A POWER SOURCE INTO THE USB PORT AND CONFIRMING THAT THE DISPLAY TURNS ON, YOU HEAR AUDIBLE BEEPS, FEEL THE PUMP VIBRATE, AND SEE THE GREEN LED LIGHT BLINKING AROUND THE EDGE OF THE SCREEN ON/QUICK BOLUS BUTTON. IF YOU ARE UNSURE ABOUT POTENTIAL DAMAGE, DISCONTINUE USE OF THE PUMP AND CONTACT CUSTOMER TECHNICAL SUPPORT.
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS INTERMITTENTLY UNRESPONSIVE. DURING TROUBLESHOOTING, THE ISSUE WAS CONFIRMED. REPORTEDLY, CUSTOMER DROPPED THE PUMP IN WATER. CUSTOMER'S BLOOD GLUCOSE WAS 190-300 MG/DL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD IN INSULIN THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2281409 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1000096 | 00853052007264 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR | Male |