FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 11.5MM

MDR report key: 15130137 · Received July 29, 2022

Report

Report Number
0002023141-2022-01918
Event Type
Injury
Date Received
July 29, 2022
Date of Event
July 18, 2022
Report Date
January 27, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K061410
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. PATIENT SEX UNKNOWN / NOT PROVIDED. WEIGHT UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER: K013227. PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2022-03100-1. THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: TYPE OF INVESTIGATION CODES WERE UPDATED: 3331, 4109, 4110, 4111. H6: INVESTIGATION FINDINGS CODE WAS UPDATED: 3252. H6: INVESTIGATION CONCLUSION CODE WAS UPDATED: 4307. ONE IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 11.5MM (TSVWB11) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED IMPLANT FRACTURED AT THE COLLAR REGION. BASED ON THE EVALUATION, DEVICE MALFUNCTION HAS OCCURRED. HOWEVER, THERE IS NO EXISTING NONCONFORMANCE / CAPA / HHE/D / IE / PRODUCT HOLD AGAINST THE REPORTED DEVICE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. MONTHLY POST MARKET TRENDING REVIEW IDENTIFIED NO ADVERSE TRENDS OR ACTIONABLE TRENDS FOR THE REPORTED EVENT OR DEVICE. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCTS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, THE PRODUCT WAS WITHIN SPECIFICATIONS AND CONFORMING WHEN IT LEFT ZIMVIE. DHR REVIEW FOR THE LOT (63164592) HAS REVEALED NO DEVIATIONS NOR NON-CONFORMANCES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENTS. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (63164592) FOR SIMILAR EVENTS. THE REPORTED EVENT COULD NOT BE RECREATED DUE TO THE NATURE OF THE DENTAL DEVICE AND EVENT. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, BASED ON THE INVESTIGATION, RISK REVIEW AND IFU, THE MOST LIKELY CAUSES DETERMINED FROM THE INVESTIGATION IS CLINICIAN¿S ERROR WHILE REMOVING IMPLANT. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND CORRECTED DATA FROM REPORT: 0002023141- 2022- 01918. CORRECTED DATA: THE PATIENT IDENTIFIER WAS CHANGED TO (B)(6). THE MANUFACTURER DATE WAS CHANGED TO AUGUST 28, 2015 AND THE EXPIRATION DATE WAS CHANGED TO AUGUST 31, 2020. THE REPORTER CONTACT INFORMATION WAS CHANGED TO DR. CIMARA FERREIRA WITH RELEVANT INFORMATION.

Description of Event or Problem · 0

CUSTOMER REPORTED A FRACTURED IMPLANT ON TOOTH #19. NO PATIENT IMPACT.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION WAS RECEIVED AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

UPON FURTHER INVESTIGATION, THE DOCTOR'S OFFICE REPORTED ADDITIONAL PATIENT INFORMATION, EVENT DATE INFORMATION, DEVICE INFORMATION AND DOCTOR CONTACT INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2654238 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL 63164592

Patients

Seq Age Sex Outcome Treatment
1 76 YR Male Required Intervention