FDA Adverse Event Malfunction Summary report: N

SPECTRUM IQ INFUSION PUMP

MDR report key: 15129640 · Received July 29, 2022

Report

Report Number
1314492-2022-03214
Event Type
Malfunction
Date Received
July 29, 2022
Date of Event
May 26, 2022
Report Date
July 29, 2022
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
UDI-DI
00085412610900
PMA / PMN Number
K220417
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RECEIVED FOR EVALUATION. DURING FUNCTIONAL TESTING, A FAILED TO ALARM UPSTREAM OCCLUSION DID NOT OCCUR. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THE REPORTED CONDITION WAS NOT VERIFIED. THERE IS CURRENTLY AN OPEN FIELD ACTION (FA-2021-056) ASSOCIATED WITH REINFORCING PROPER IV LINE SETUP AND DETECTION OF UPSTREAM OCCLUSION FOR SPECTRUM PUMPS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SPECTRUM IQ PUMP FAILED TO ALARM UPSTREAM OCCLUSION DURING THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2426218 SPECTRUM IQ INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION NA NA 00085412610900

Patients

Seq Age Sex Outcome Treatment
1 Unknown