BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD
Report
- Report Number
- 9680577-2022-00085
- Event Type
- Malfunction
- Date Received
- July 28, 2022
- Date of Event
- July 5, 2022
- Report Date
- August 10, 2022
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- JSJ
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6 INVESTIGATION SUMMARY : THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR RECENT COMPLAINT (B)(4) AGAINST COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD, CATALOG NUMBER 254007, LOT NUMBER 2129287. EVENT DESCRIPTION: THE CUSTOMER REPORTED SEVERAL PLATES BEING NOT SELECTIVE FOR GRAM POSITIVE BACTERIA, AS GROWTH, SOMETIMES EVEN CONSIDERABLE, OF GRAM NEGATIVE BACTERIA HAS OCCURRED ON THEM. COMPLAINT HISTORY REVIEW: THE COMPLAINT HISTORY HAS BEEN REVIEWED FOR A PERIOD OF 12 MONTHS AND NO SIMILAR ISSUE WAS REPORTED FOR THIS CATALOG NUMBER. BATCH HISTORY RECORD (BHR) REVIEW: THE BATCH HISTORY HAS BEEN REVIEWED AND NO DEVIATION WAS RECORDED. ALL THE RELEASE TESTING WAS SATISFACTORY. SAMPLE ANALYSIS: A PERFORMANCE TEST HAS BEEN CONDUCTED ON THE RETAINING SAMPLES FOR THE STRAINS THAT SHOULD BE INHIBITED (ATCC 27853 ; PS. AERUGINOSA AND ATCC 12453; PROT. MIRABILIS) WITH EXCELLENT RESULTS. THE STRAINS WERE INHIBITED. RETURN SAMPLES WERE NOT PROVIDED HOWEVER PICTURE ILLUSTRATING A PLATE WITH BACTERIAL GROWTH WAS PROVIDED. EVALUATION RESULTS: BASED UPON OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMATIC FAILURE IN OUR MANUFACTURING PROCESS. ALL THE RELEASE TESTING WAS SATISFACTORY AND ALSO THE RETEST PERFORMED ON THE RETAIN SAMPLES WITH ATCC 27853 (PS. AERUGINOSA) AND ATCC 12453 (PROT. MIRABILIS) DID NOT REVEAL ANY DEVIATION TO THE REQUIREMENT. SINCE NO TREND WAS IDENTIFIED AND NO PROCESS DEVIATION COULD BE DETECTED, NO ACTION WILL BE CONDUCTED THIS TIME. INVESTIGATION CONCLUSION: BASED ON THE ABOVE MENTIONED INVESTIGATIONS AND TESTING, WE CANNOT CONFIRM THIS COMPLAINT. HOWEVER BD WILL CONTINUE MONITORING INCOMING COMPLAINT WITH SIMILAR FAILURE MODE. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT WHILE USING BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD ATYPICAL GROWTH WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SEVERAL PLATES ARE NOT SELECTIVE FOR GRAM POSITIVE BACTERIA, AS GROWTH, SOMETIMES EVEN CONSIDERABLE, OF GRAM NEGATIVE BACTERIA HAS OCCURRED ON THEM."
IT WAS REPORTED THAT WHILE USING BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD ATYPICAL GROWTH WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SEVERAL PLATES ARE NOT SELECTIVE FOR GRAM POSITIVE BACTERIA, AS GROWTH, SOMETIMES EVEN CONSIDERABLE, OF GRAM NEGATIVE BACTERIA HAS OCCURRED ON THEM."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2924523 | BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD | CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL | JSJ | BECTON DICKINSON GMBH | 2129287 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |