STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2022-12596
- Event Type
- Injury
- Date Received
- July 28, 2022
- Date of Event
- June 15, 2022
- Report Date
- August 30, 2022
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(6). A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE REASON FOR REOPERATION IS: "DEFLATION".
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION, RECEIVED ON JUL 29, 2022 WITH LOT NUMBER 1032751. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: WEAR ABRASION, AN OPENING ON VALVE BOND EDGE, AND CREASE FLAT. LEAK TEST AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED: A SHARP OPENING ON VALVE BOND EDGE. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFIED THE THICKNESS WITHIN SPECIFICATION. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: - A SHARP OPENING ON VALVE BOND EDGE ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2287369 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 1032751 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Female | Required Intervention |