FDA Adverse Event Malfunction Summary report: N

CARDIOCHEK PLUS ANALYZER

MDR report key: 15121718 · Received July 28, 2022

Report

Report Number
1836135-2022-00037
Event Type
Malfunction
Date Received
July 28, 2022
Date of Event
June 28, 2022
Report Date
July 28, 2022
Manufacturer
PTS DIAGNOSTICS
Product Code
CGA
UDI-DI
00381932700016
PMA / PMN Number
K140068
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DURING THE INVESTIGATION, IT IS DETERMINED THAT THE TERMINAL DISPLACEMENT IS MOST LIKELY CAUSED BY THE CUSTOMER PROVIDING UNDUE FORCE ONTO THE BATTERY TERMINALS. IN ADDITION, THE POTENTIAL EXISTS FOR THE BATTERY CONTACT TO LOSE TRACTION WITH THE COMPARTMENT CAUSING LESS FORCE TO DISLODGE HOWEVER IMPROPER REMOVAL OF BATTERIES WILL STILL FORCE THE DISLODGE. SEVERAL KEY FACTORS POINT TO THIS DECISION: 1. THE BATTERY TERMINALS WOULD HAVE TO HAVE BEEN PROPERLY SEATED AT THE TIME OF MANUFACTURING BECAUSE THE ANALYZER WOULD NOT POWER ON AND WOULD CONSEQUENTLY BE UNABLE TO BE CALIBRATED AND RELEASED. 2. UPON CLOSE INSPECTION OF THE BATTERY TERMINAL AREA, IT WAS OBSERVED THAT THE TERMINALS WERE FULLY SEATED INITIALLY DUE TO VISIBLE DRAG MARKS CREATED BY THE METAL TERMINAL SCRAPING THE PLASTIC CASING WHEN EXCESSIVE FORCE WAS APPLIED. THESE MARKINGS ARE PRESENT FROM THE POINT OF TERMINAL SEATING DOWN TO WHERE THE TERMINALS LIE CURRENTLY. PTS IS CONFIDENT THAT THIS DAMAGE OCCURRED POST-DISTRIBUTION, AS THERE ARE NO MANUFACTURING STEPS THAT WOULD CAUSE THIS TYPE OF DAMAGE. THE CUSTOMER ALLEGATION OF OVERHEATING WAS NOT OBSERVED ON THE RETURNED ANALYZER. NUMEROUS BATTERY COMPARTMENT IMPROVEMENTS HAVE ALREADY BEEN IMPLEMENTED PRIOR TO THESE ALLEGATIONS FROM THE FIELD. PROCESS IMPROVEMENTS AND PART SPECIFICATIONS HAVE ENHANCED SEATING AND HOLD CAPACITY OF THE BATTERY CONTACTS. ADDITIONALLY, A NEW BATTERY COMPARTMENT DESIGN (APPROVED BY THE FDA PER K193406) ELIMINATES THE POTENTIAL FOR OVERHEATING BY REMOVING THE POINT OF CONTACT THAT AN UNSEATED TERMINAL WOULD REACH.

Description of Event or Problem · 0

INITIAL COMMUNICATION FROM THE CUSTOMER STATED THAT AFTER CHANGING BATTERIES ON THEIR ANALYZER IT WOULD NOT POWER ON. UPON FOLLOW-UP FOR TROUBLESHOOTING, THE CUSTOMER NOTED THAT THE ANALYZER ALSO "GOT EXTREMELY HOT" AT ONE POINT. THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2320957 CARDIOCHEK PLUS ANALYZER CLINICAL CHEMISTRY ANALYZER CGA PTS DIAGNOSTICS 2700 00381932700016

Patients

Seq Age Sex Outcome Treatment
1 Unknown