FDA Adverse Event Malfunction Summary report: N

WATER FILTER

MDR report key: 15116713 · Received July 27, 2022

Report

Report Number
9610595-2022-00354
Event Type
Malfunction
Date Received
July 27, 2022
Report Date
October 7, 2022
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

FULL NAME OF THE FACILITY IS (B)(6) HOSPITAL. THERE ARE TOTAL 9 DEVICES ASSOCIATED WITH THIS EVENT (FOUR AUTOMATIC ENDOSCOPY REPROCESSORS (AERS), FOUR WATER FILTERS, AND ONE UNKNOWN SCOPE). THESE DEVICES ARE CAPTURED IN MEDWATCHES WITH PATIENT IDENTIFIERS AS BELOW: PATIENT IDENTIFIER , MODEL #, SERIAL #. (B)(6), OER-5 (AER), 2101296. (B)(6), OER-5 (AER), 2101297. (B)(6), OER-6 (AER), 2100681. (B)(6), OER-6 (AER), 2100686. (B)(6), MAJ-2317, (WATER FILTER). (B)(6), MAJ-2317, (WATER FILTER). (B)(6), MAJ-2317, (WATER FILTER). (B)(6), MAJ-2317, (WATER FILTER). (B)(6), CF-H290I, REPRESENTING THE UNKNOWN SCOPE. THIS MEDWATCH IS FOR THE PATIENT IDENTIFIER (B)(6). THE DEVICE HAS BEEN RETURNED BUT THE DEVICE EVALUATION IS NOT YET COMPLETED. AS SUCH, A DEFINITIVE ROOT CAUSE OF THE REPORTED COMPLAINT CANNOT BE DETERMINED AT THIS TIME. SUPPLEMENTAL REPORT(S) WILL BE FILED AS ANY RELEVANT NEW INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE APPROVED FINAL INVESTIGATION AND DEVICE EVALUATION. THE DEVICE HISTORY RECORD WAS UNABLE TO BE REVIEWED FOR THIS DEVICE SINCE THE SERIAL NUMBER WAS NOT PROVIDED. HOWEVER, OLYMPUS ONLY RELEASES PRODUCTS TO MARKET THAT MEET ALL MANUFACTURING SPECIFICATIONS AND FINAL PRODUCT RELEASE CRITERIA. BASED ON THE RESULTS OF THE INVESTIGATION, THE DIRTY WATER HAD ACCUMULATED INSIDE OF THE WATER FILTER, THE WATER FILTER TURNED BROWN, AND HAD A FAINT, STRANGE SMELL. THE LIKELY ROOT CAUSE OF THE DIRTY WATER WAS POOR REPROCESSING. IT IS POSSIBLE THAT THE WATER FILTER CLOGGED BECAUSE IT WAS NOT REPLACED, AND THAT WATER WAS NOT PROPERLY REMOVED. AS A RESULT, IT IS POSSIBLE SUFFICIENT REPROCESSING COULD NOT BE PERFORMED AND THE CONCENTRATION OF ACECIDE DECREASED. THE INSTRUCTION MANUAL IDENTIFIES THE FOLLOWING VERBIAGE, WHICH MAY HAVE PREVENTED THE PHENOMENON: ¿REPLACE THE WATER FILTER PERIODICALLY, AT LEAST ONCE IN TWO MONTHS TO PREVENT CONTAMINATION OF THE RINSE WATER. REPLACE THE WATER FILTER AT LEAST ONCE IN TWO MONTHS. OTHERWISE, MISCELLANEOUS GERMS AND STAINS IN THE WATER MAY CONTAMINATE THE EQUIPMENT AND/OR THE ENDOSCOPES AND PREVENT EFFECTIVE REPROCESSING. DISINFECTING THE WATER SUPPLY CHANNEL REDUCES THE CONCENTRATION OF THE DISINFECTANT SOLUTION, MAKING IT UNUSABLE. THEREFORE, YOU ARE RECOMMENDED TO DISINFECT THE WATER SUPPLY CHANNEL IMMEDIATELY BEFORE REPLACING THE DISINFECTANT SOLUTION.¿ OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

AS REPORTED FOR THIS EVENT BY THE FACILITY, THE DEVICE (WATER FILTER) OF AUTOMATIC ENDOSCOPY REPROCESSOR (AER) HAD NOT BEEN REPLACED SINCE THE FIRST USE OF THE AER DEVICE APPROXIMATELY EIGHT MONTHS BEFORE EVENT WAS REPORTED. THERE IS NO HARM REPORTED TO ANY PATIENT OR PERSONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2440103 WATER FILTER WATER FILTER FEB AIZU OLYMPUS CO., LTD. MAJ-2317

Patients

Seq Age Sex Outcome Treatment
1 Unknown