FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 15113727 · Received July 27, 2022

Report

Report Number
9617229-2022-12436
Event Type
Injury
Date Received
July 27, 2022
Date of Event
June 9, 2022
Report Date
July 27, 2022
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION. DHR TREND SUMMARY REVIEW OF ALL COMPLAINTS OF A050401 - FLUID LEAK FOR THE PERIOD OF JUN 2020 THROUGH MAY 2022 RELATED TO DATE OPENED INDICATES THAT 1 POINT IS ABOVE THE UPPER CONTROL LIMIT. ACCORDING TO THE SEVERITY, AN ADDITIONAL ANALYSIS WAS PERFORMED TO REVIEW THE INVOLVED COMPLAINTS WITH MANUFACTURING DATE OVER THE PAST 24 MONTHS. SEE ¿I CHART OF A050401 FLUID LEAK FOR SALINE BREAST IMPLANTS¿ ATTACHED. NO PATTERNS WERE FOUND; THEREFORE, NO ADDITIONAL ACTIONS ARE DEEMED REQUIRED. THE TREND OF A050401 - FLUID LEAK COMPLAINTS WILL CONTINUE TO BE MONITORED AND CORRECTIVE ACTIONS WILL BE TAKEN IN THE FUTURE IF DEEMED APPROPRIATE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1703991 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 1421051

Patients

Seq Age Sex Outcome Treatment
1 35 YR Female Required Intervention