FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 15113575 · Received July 27, 2022

Report

Report Number
3013756811-2022-77136
Event Type
Malfunction
Date Received
July 27, 2022
Date of Event
July 7, 2022
Report Date
July 7, 2022
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE NOT RETURNED.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN ALTITUDE ALARM OCCURRED WHILE THE CUSTOMER WAS NOT OUTSIDE OF THE LABELED OPERATING ALTITUDE RANGE. REPORTEDLY, CUSTOMER CLEANED THE SPEAKER HOLES AND CLEARED THE ALARM. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 180-307 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2137856 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717

Patients

Seq Age Sex Outcome Treatment
1 18 YR Female