FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ 3-PIECE SYRINGE

MDR report key: 15112180 · Received July 27, 2022

Report

Report Number
3003152976-2022-00334
Event Type
Malfunction
Date Received
July 27, 2022
Date of Event
June 20, 2022
Report Date
August 22, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY ONE SAMPLE AND SEVERAL PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, A FOREIGN MATTER WAS OBSERVED IN THE PLUNGER. FURTHER EVALUATION IDENTIFIED THE MATTER TO BE GREASE, CONSISTENT WITH WHAT IS USED DURING THE MOLDING PROCESS. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2103040, NO DEVIATIONS OR NON-CONFORMANCES RELATED TO THE REPORTED ISSUE WERE IDENTIFIED DURING THE MANUFACTURING PROCESS. WHILE WE COULD NOT IDENTIFY A DIRECT ISSUE, THIS DEFECT CAN OCCUR DUE TO GREASE ACCUMULATING IN THE MOLD DURING THE MOLDING PROCESS. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHITE FOREIGN MATTER WAS FOUND INSIDE EIGHT BD PLASTIPAK¿ 3-PIECE SYRINGE WHILE PREPARING MEDICATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DURING MEDICATION PREPARATION A NURSE DISCOVERED THAT THERE WAS SOMETHING INSIDE THE BD PLASTIPAK SYRINGE. SHE HAD ONLY MANAGED TO PULL THE PLUNGER BACK A LITTLE TO FILL THE SYRINGE WITH AIR WHEN SHE DISCOVERED THIS. THE MATERIAL INSIDE THE SYRINGE LOOKS WHITE AND THREAD LIKE, THE MATERIAL IS ON THE PLUNGER, SLIGHTLY ON THE SIDE OF THE PLUNGER AND AT THE VERY TOP OF THE SYRINGE. THE NURSE FOUND SOME OTHER BD PLASTIPAK WITH THE SAME LOT AT THEIR LOCAL STORAGE, BUT THEY DO NOT SEEM TO HAVE THE SAME MATERIAL ON THEM."

Description of Event or Problem · 0

IT WAS REPORTED THAT WHITE FOREIGN MATTER WAS FOUND INSIDE EIGHT BD PLASTIPAK¿ 3-PIECE SYRINGE WHILE PREPARING MEDICATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DURING MEDICATION PREPARATION A NURSE DISCOVERED THAT THERE WAS SOMETHING INSIDE THE BD PLASTIPAK SYRINGE. SHE HAD ONLY MANAGED TO PULL THE PLUNGER BACK A LITTLE TO FILL THE SYRINGE WITH AIR WHEN SHE DISCOVERED THIS. THE MATERIAL INSIDE THE SYRINGE LOOKS WHITE AND THREAD LIKE, THE MATERIAL IS ON THE PLUNGER, SLIGHTLY ON THE SIDE OF THE PLUNGER AND AT THE VERY TOP OF THE SYRINGE. THE NURSE FOUND SOME OTHER BD PLASTIPAK WITH THE SAME LOT AT THEIR LOCAL STORAGE, BUT THEY DO NOT SEEM TO HAVE THE SAME MATERIAL ON THEM."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722936 BD PLASTIPAK¿ 3-PIECE SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2103040

Patients

Seq Age Sex Outcome Treatment
1 Unknown