FDA Adverse Event Malfunction Summary report: N

ZIMMER AIR DERMATOME

MDR report key: 15111811 · Received July 27, 2022

Report

Report Number
0001526350-2022-00704
Event Type
Malfunction
Date Received
July 27, 2022
Date of Event
February 18, 2022
Report Date
July 27, 2022
Manufacturer
ZIMMER SURGICAL, INC.
Product Code
GFD
PMA / PMN Number
EXEMPT
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS COMPLAINT IS RECORDED BY ZIMMER BIOMET UNDER (B)(4). REVIEW OF THE MOST RECENT REPAIR RECORD IDENTIFIED THE FOLLOWING RELATED REPAIR: TECH CONFIRMED THE REPORTED ISSUE AS THE MOTOR HAD SEIZED DURING ASSESSMENT. ADDITIONALLY THE TECH FOUND THAT THE MACHINED HEAD WAS DAMAGED AND SCREWS WERE WORN, THE UNIT WAS OUT OF CALIBRATION, AND THE CONTROL BAR WAS OUT OF POSITION. TECH REPLACED THE MOTOR, MOTOR SLEEVE, MACHINED HEAD, SHAFT BEARINGS, SLEEVE BEARINGS, SCREWS, AND THEN REPOSITIONED THE CONTROL BAR. TECH TESTED THE UNIT AND RETURNED IT TO SERVICE. THE EVENT IS CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

FOLLOWING SECTIONS WERE UPDATED OR CORRECTED : G3 IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS

Description of Event or Problem · 0

IT WAS REPORTED THAT THE HANDPIECE GETS STUCK. DURING INITIAL EVALUATION OF THE DEVICE, THE MOTOR EXHIBITED INCONSISTENT SPEED INCLUDING SEIZING. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
706902 ZIMMER AIR DERMATOME DERMATOME, PNEUMATICALLY-POWERED GFD ZIMMER SURGICAL, INC. N/A 60223471

Patients

Seq Age Sex Outcome Treatment
1 Unknown