ZIMMER AIR DERMATOME
Report
- Report Number
- 0001526350-2022-00704
- Event Type
- Malfunction
- Date Received
- July 27, 2022
- Date of Event
- February 18, 2022
- Report Date
- July 27, 2022
- Manufacturer
- ZIMMER SURGICAL, INC.
- Product Code
- GFD
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- 003
Narratives
THIS COMPLAINT IS RECORDED BY ZIMMER BIOMET UNDER (B)(4). REVIEW OF THE MOST RECENT REPAIR RECORD IDENTIFIED THE FOLLOWING RELATED REPAIR: TECH CONFIRMED THE REPORTED ISSUE AS THE MOTOR HAD SEIZED DURING ASSESSMENT. ADDITIONALLY THE TECH FOUND THAT THE MACHINED HEAD WAS DAMAGED AND SCREWS WERE WORN, THE UNIT WAS OUT OF CALIBRATION, AND THE CONTROL BAR WAS OUT OF POSITION. TECH REPLACED THE MOTOR, MOTOR SLEEVE, MACHINED HEAD, SHAFT BEARINGS, SLEEVE BEARINGS, SCREWS, AND THEN REPOSITIONED THE CONTROL BAR. TECH TESTED THE UNIT AND RETURNED IT TO SERVICE. THE EVENT IS CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
FOLLOWING SECTIONS WERE UPDATED OR CORRECTED : G3 IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS
IT WAS REPORTED THAT THE HANDPIECE GETS STUCK. DURING INITIAL EVALUATION OF THE DEVICE, THE MOTOR EXHIBITED INCONSISTENT SPEED INCLUDING SEIZING. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.
NO ADDITIONAL EVENT INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 706902 | ZIMMER AIR DERMATOME | DERMATOME, PNEUMATICALLY-POWERED | GFD | ZIMMER SURGICAL, INC. | N/A | 60223471 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |