CLEARGLIDE EVH SMALL
Report
- Report Number
- 1718850-2009-00073
- Event Type
- Malfunction
- Date Received
- October 14, 2009
- Date of Event
- September 14, 2009
- Report Date
- September 14, 2009
- Manufacturer
- SORIN GROUP USA
- Product Code
- GEI
- PMA / PMN Number
- K003587
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
THE 510(K) NUMBER FOR THE BIPOLAR DEVICE IS K003587. THIS DEVICE IS A COMPONENT IN THE CLEARGLIDE EVH SMALL KTV15 KIT. THE CLINICIAN REPORTED THAT THE JAW TIP BROKE OFF THE ENDOSCOPIC VEIN HARVESTING BIPOLAR DEVICE DURING CLEANING. THEREFORE, THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT INJURY. ONE BIPOLAR DEVICE WAS RECEIVED FOR EVALUATION. THE VISUAL INSPECTION CONFIRMED THAT THE TIP WAS BROKEN. AFTER INSPECTION AT SORIN GROUP USA, THE DEVICE WAS SENT TO THE VENDOR FOR FURTHER ANALYSIS. A FOLLOW-UP REPORT WILL BE FILED WHEN THE EVALUATION IS COMPLETE. THE BIPOLAR INSTRUCTIONS FOR USE STATE "DO NOT ADVANCE OR TORQUE THE INSTRUMENT WITH THE JAWS OPEN OR USE THE JAWS FOR SPREAD DISSECTION. THESE ACTIONS WILL DAMAGE THE INSTRUMENT."
THE CLINICIAN REPORTED THAT THE JAW TIP BROKE OFF THE ENDOSCOPIC VEIN HARVESTING BIPOLAR DEVICE WHILE CLEANING IT. THE INCIDENT OCCURRED DURING CLEANING. THEREFORE, THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CLEARGLIDE EVH SMALL | BIPOLAR DEVICE | GEI | SORIN GROUP USA | NA | 0921500031 |