FDA Adverse Event
Injury
Summary report: N
ROCHE ACCUCHEK INFORM METERS
MDR report key: 1510278
·
Received September 28, 2009
Report
- Report Number
- MW5013023
- Event Type
- Injury
- Date Received
- September 28, 2009
- Date of Event
- September 12, 2009
- Report Date
- September 25, 2009
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
MD FINAL SUMMARY: PT ON PERITONEAL DIALYSIS AND PLAN WAS FOR DISCHARGE BUT THEN, PT HAD SEIZURE. APPARENTLY, MALTOSE IN HIS DIALYSATE WAS INTERFERING WITH HOSPITAL GLUCOMETER AND HE WAS BECOMING HYPOGLYCEMIC THOUGH FSG (FINGER STICK GLUCOSE) REMAINED NORMAL. TREATED WITH DILANTIN AND APPROPRIATE GLYCEMIC CONTROL WITH NO FURTHER SEIZURES. REQUIRED TRANSFER TO HIGHER CARE LEVEL TO STABILIZE GLYCEMIC CONTROL AND LOS EXTENDED BY 7 DAYS. DATES OF USE: 2009. DIAGNOSIS OR REASON FOR USE: TYPE II DM FOR GLYCEMIC CTRL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ROCHE ACCUCHEK INFORM METERS | GLUCOMETER | NBW | ROCHE DIAGNOSTICS | 2001201 | 550925 | |
| 2 | ACCU-CHEK COMFORT CURVE TEST STRIPS | NONE | LFR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention |