FDA Adverse Event Injury Summary report: N

ROCHE ACCUCHEK INFORM METERS

MDR report key: 1510278 · Received September 28, 2009

Report

Report Number
MW5013023
Event Type
Injury
Date Received
September 28, 2009
Date of Event
September 12, 2009
Report Date
September 25, 2009
Manufacturer
ROCHE DIAGNOSTICS
Product Code
NBW
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

MD FINAL SUMMARY: PT ON PERITONEAL DIALYSIS AND PLAN WAS FOR DISCHARGE BUT THEN, PT HAD SEIZURE. APPARENTLY, MALTOSE IN HIS DIALYSATE WAS INTERFERING WITH HOSPITAL GLUCOMETER AND HE WAS BECOMING HYPOGLYCEMIC THOUGH FSG (FINGER STICK GLUCOSE) REMAINED NORMAL. TREATED WITH DILANTIN AND APPROPRIATE GLYCEMIC CONTROL WITH NO FURTHER SEIZURES. REQUIRED TRANSFER TO HIGHER CARE LEVEL TO STABILIZE GLYCEMIC CONTROL AND LOS EXTENDED BY 7 DAYS. DATES OF USE: 2009. DIAGNOSIS OR REASON FOR USE: TYPE II DM FOR GLYCEMIC CTRL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ROCHE ACCUCHEK INFORM METERS GLUCOMETER NBW ROCHE DIAGNOSTICS 2001201 550925
2 ACCU-CHEK COMFORT CURVE TEST STRIPS NONE LFR

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention