RESIUS 190 XPRES
Report
- Report Number
- 1509706
- Event Type
- Malfunction
- Date Received
- October 13, 2009
- Date of Event
- October 12, 2009
- Report Date
- October 12, 2009
- Manufacturer
- KONICA-MINOLTA MEDICAL IMAGING USA, INC.
- Product Code
- IWZ
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CT, US
- Reporter Occupation
- RISK MANAGER
Narratives
OF THE 108 PLATES PUT INTO USE, 60 PLATES HAVE DEVELOPED FLAWS OR ARTIFACTS. THESE ARTIFACTS IF NOT DETECTED COULD RESULT IN ADDITIONAL DIAGNOSTIC TESTING AND POTENTIAL HARM.BEGINNING IN SEPT 2007, THIS FACILITY RETURNED 28 OF THE 43 14" X 17" PLATES UNDER WARRANTY TO KONICA AND RECEIVED REPLACEMENT. SUBSEQUENTLY, AN ADDITIONAL 41 PLATES WERE FOUND TO BE FLAWED. KONICA TOLD THIS HOSPITAL THAT NO OTHER FACILITIES HAVE REPORTED THIS TYPE OF ISSUE.THE ARTIFACT ON THE IMAGING PLATES APPEARS AS A SMALL AREA(S) OF HIGHER ATTENTUATION THAT CAN BE MISTAKEN FOR CALCIFICATION.WE BELIEVE THAT THE CAUSE MAY BE CHIPS OF GLUE WHICH FALL BETWEEN THE COVER AND THE IMAGED STORAGE PLATE. THE GLUE IS USED TO HOLD THE METAL FRAME SURROUNDING THE CASSETTE.
OF THE 108 PLATES PUT INTO USE, 60 PLATES HAVE DEVELOPED FLAWS OR ARTIFACTS. THESE ARTIFACTS IF NOT DETECTED COULD RESULT IN ADDITIONAL DIAGNOSTIC TESTING AND POTENTIAL HARM.BEGINNING IN SEPT 2007, THIS FACILITY RETURNED 28 OF THE 43 14" X 17" PLATES UNDER WARRANTY TO KONICA AND RECEIVED REPLACEMENT. SUBSEQUENTLY, AN ADDITIONAL 41 PLATES WERE FOUND TO BE FLAWED. KONICA TOLD THIS HOSPITAL THAT NO OTHER FACILITIES HAVE REPORTED THIS TYPE OF ISSUE.THE ARTIFACT ON THE IMAGING PLATES APPEARS AS A SMALL AREA(S) OF HIGHER ATTENTUATION THAT CAN BE MISTAKEN FOR CALCIFICATION.WE BELIEVE THAT THE CAUSE MAY BE CHIPS OF GLUE WHICH FALL BETWEEN THE COVER AND THE IMAGED STORAGE PLATE. THE GLUE IS USED TO HOLD THE METAL FRAME SURROUNDING THE CASSETTE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESIUS 190 XPRES | IMAGING PLATE, COMPUTED RADIOGRAPHY | IWZ | KONICA-MINOLTA MEDICAL IMAGING USA, INC. | RESIUS 190 XPRESS | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |