FDA Adverse Event Malfunction Summary report: N

CMOS VIDEO CYSTO-URETHROSCOPE C-VIEW REV DEF

MDR report key: 15093886 · Received July 25, 2022

Report

Report Number
1221826-2022-00128
Event Type
Malfunction
Date Received
July 25, 2022
Date of Event
June 27, 2022
Report Date
November 10, 2022
Manufacturer
KARL STORZ ENDOVISION, INC.
Product Code
FAJ
PMA / PMN Number
K202957
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PER KSE/MFR. EVALUATION IS PENDING.

Additional Manufacturer Narrative · 0

IMAGE INTERRUPTION DUE TO ESU (ELECTROSURGICAL UNIT) FAILURE. EXTERNAL CUT ON SHAFT BEHIND PROXIMAL THREAD WRAP, LEAKS. OVER BEND ON PROXIMAL SHAFT 65MM FROM SHAFT ADAPTOR. LIGHT GUIDE PULLED BACK FROM DISTAL HEAD CAUSING LOW LIGHT OUTPUT, 30MW. OVER TEMPERATURE DOT TRIGGERED AT 180°F. RISK MANAGEMENT FILE HAS BEEN REVIEWED AND THIS HAZARD IS COVERED. THIS FAILURE MODE WAS CONFIRMED AND HAS BEEN DETERMINED TO BE DUE TO THE DURABILITY OF THE POWER CABLE. IAF 2022-07-25-01 HAS BEEN FILED FOR CAPA CONSIDERATION. CAPA 2022-0111 HAS BEEN ENTERED.

Description of Event or Problem · 0

IMAGE INTERRUPTION, THE MONITOR KEPT SHORTING OUT AND RESTARTING WHEN USING CAUTERY. THE PROCEDURE WAS COMPLETED. THE FACILITY PROVIDED NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1411471 CMOS VIDEO CYSTO-URETHROSCOPE C-VIEW REV DEF FLEXIBLE VIDEO CYSTO SCOPE FAJ KARL STORZ ENDOVISION, INC. R11272VUEK

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other