FDA Adverse Event
Malfunction
Summary report: N
CMOS VIDEO CYSTO-URETHROSCOPE C-VIEW REV DEF
MDR report key: 15093886
·
Received July 25, 2022
Report
- Report Number
- 1221826-2022-00128
- Event Type
- Malfunction
- Date Received
- July 25, 2022
- Date of Event
- June 27, 2022
- Report Date
- November 10, 2022
- Manufacturer
- KARL STORZ ENDOVISION, INC.
- Product Code
- FAJ
- PMA / PMN Number
- K202957
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
PER KSE/MFR. EVALUATION IS PENDING.
Additional Manufacturer Narrative · 0
IMAGE INTERRUPTION DUE TO ESU (ELECTROSURGICAL UNIT) FAILURE. EXTERNAL CUT ON SHAFT BEHIND PROXIMAL THREAD WRAP, LEAKS. OVER BEND ON PROXIMAL SHAFT 65MM FROM SHAFT ADAPTOR. LIGHT GUIDE PULLED BACK FROM DISTAL HEAD CAUSING LOW LIGHT OUTPUT, 30MW. OVER TEMPERATURE DOT TRIGGERED AT 180°F. RISK MANAGEMENT FILE HAS BEEN REVIEWED AND THIS HAZARD IS COVERED. THIS FAILURE MODE WAS CONFIRMED AND HAS BEEN DETERMINED TO BE DUE TO THE DURABILITY OF THE POWER CABLE. IAF 2022-07-25-01 HAS BEEN FILED FOR CAPA CONSIDERATION. CAPA 2022-0111 HAS BEEN ENTERED.
Description of Event or Problem · 0
IMAGE INTERRUPTION, THE MONITOR KEPT SHORTING OUT AND RESTARTING WHEN USING CAUTERY. THE PROCEDURE WAS COMPLETED. THE FACILITY PROVIDED NO ADDITIONAL INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1411471 | CMOS VIDEO CYSTO-URETHROSCOPE C-VIEW REV DEF | FLEXIBLE VIDEO CYSTO SCOPE | FAJ | KARL STORZ ENDOVISION, INC. | R11272VUEK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |