FDA Adverse Event Malfunction Summary report: N

MYCOPHENOLIC ACID

MDR report key: 15089564 · Received July 22, 2022

Report

Report Number
1823260-2022-02172
Event Type
Malfunction
Date Received
July 22, 2022
Date of Event
June 24, 2022
Report Date
September 14, 2022
Manufacturer
ROCHE DIAGNOSTICS
Product Code
OAV
UDI-DI
04015630997381
PMA / PMN Number
K063520
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

QUALITY CONTROLS ARE WITHIN RANGE EXCEPT THE FIRST LEVEL OF CONTROL DRIFTS UPWARD. NO LEVEL 1 CONTROL INFORMATION WAS AVAILABLE FOR THE TIME OF THE INCIDENT. THE LAST 10 CALIBRATIONS WERE STABLE. THE PATIENT'S FLUCONAZOLE MEDICATION DOES NOT INTERFERE WITH THE ASSAY AND INTERFERENCE WITH THE PIRACETAM MEDICATION IS UNLIKELY. THE INVESTIGATION DETERMINED THAT THE ISSUE IS CONSISTENT WITH HIGH DIURESIS AND HIGH AMOUNTS OF HERBAL INFUSION. A PRODUCT PROBLEM CAN MOST LIKELY BE EXCLUDED.

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THEY RECEIVED A QUESTIONABLE RESULT FOR ONE PATIENT SAMPLE TESTED WITH MPA MYCOPHENOLIC ACID ON A COBAS 6000 C (501) MODULE (SERIAL NUMBER (B)(4). THE SAMPLE VALUE DID NOT CORRESPOND TO THE PATIENT'S CONDITION. THE SAMPLE RESULTED IN A MYCOPHENOLIC ACID VALUE OF 2.37 UG/ML WHEN TESTED ON THE C501 ANALYZER. THE VALUE WAS REPORTED OUTSIDE OF THE LABORATORY TO THE PATIENT'S DOCTOR. THE VALUE DID NOT AGREE WITH THE PATIENT'S CONDITION. THE SAMPLE WAS SENT TO A SECOND SITE WHERE IT WAS TESTED USING AN UNKNOWN METHOD ON A SIEMENS ANALYZER AND THE RESULT WAS SIMILAR TO THE ONE OBTAINED ON THE C501 ANALYZER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
546968 MYCOPHENOLIC ACID MYCOPHENOLIC ACID TEST SYSTEM OAV ROCHE DIAGNOSTICS NA 570241 04015630997381

Patients

Seq Age Sex Outcome Treatment
1 Male CELLCEPT| FLUCONAZOLE (3 TIMES WEEKLY)| HERBAL INFUSIONS| PIRACETAM (TWICE DAILY)