FDA Adverse Event Malfunction Summary report: N

FUSION OMNI-TOME

MDR report key: 15089187 · Received July 22, 2022

Report

Report Number
1037905-2022-00408
Event Type
Malfunction
Date Received
July 22, 2022
Date of Event
May 13, 2022
Report Date
July 22, 2022
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
KNS
UDI-DI
10827002319032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PMA/510(K) # K172288. INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE CUTTING WIRE SECURING COMPONENT LOCATED NEAR THE DISTAL END OF THE SPHINCTEROTOME HAS DISCONNECTED FROM THE CATHETER. THE CUTTING WIRE IS INTACT AND REMAINS SECURELY ATTACHED TO THE SPHINCTEROTOME AT THE PROXIMAL END. HOWEVER, DUE TO THE CATHETER AND SECURING COMPONENT DISCONNECTION, THE DISTAL END OF THE CUTTING WIRE IS NO LONGER CONNECTED TO THE CATHETER. THE SECURING COMPONENT HAS A LONGER SECTION THAT IS ESTIMATED TO BE 3 MM IN LENGTH AND 0.5 MM IN DIAMETER THAT HAS DETACHED FROM THE DISTAL END OF THE CUTTING WIRE AND WAS NOT INCLUDED IN THE RETURN. THE SHORTER SECTION OF THE ANCHOR IS ATTACHED TO THE TIP OF THE CUTTING WIRE, AND IT MEASURES 2 MM. DURING A CLOSE VISUAL INSPECTION OF THE DISTAL END, EVIDENCE OF A CAUTERY APPLICATION WAS OBSERVED ON THE CUTTING WIRE (BLACKENING OF THE CUTTING WIRE WAS NOTED). THE CUTTING WIRE LUMEN HAS ALSO SPLIT 13 MM PROXIMALLY. THE CANNULA (SMALL METAL TUBE) THAT RESTS INSIDE THE CUTTING WIRE LUMEN HAS ALSO MOVED AND IS RESTING ON THE CUTTING WIRE OUTSIDE OF THE CATHETER. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A VISUAL INSPECTION OF THE DEVICE CONFIRMED THE CUTTING WIRE SECURING COMPONENT (ANCHOR) HAS SEPARATED FROM CATHETER (W/DETACHMENT). A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER CAN OCCUR IF THE TIP OF THE SPHINCTEROTOME IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER WITH THE FOLLOWING COMMENT: ¿DO NOT OVER FLEX OR BOW THE TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE OR CAUSE THE CUTTING WIRE TO BREAK.¿ OTHER FACTORS THAT CAN CONTRIBUTE TO SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING PRECAUTION: "DO NOT EXERCISE THE HANDLE WHILE THE DEVICE IS COILED OR THE PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO THE SPHINCTEROTOME AND RENDER IT INOPERABLE." THE INSTRUCTIONS FOR USE ADVISE THE USER WITH THE FOLLOWING STATEMENT: ¿UPON REMOVING THE DEVICE FROM THE PACKAGE, UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME. CAREFULLY REMOVE THE PRECURVED STYLET FROM THE CANNULATING TIP.¿ PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. THE FUNCTIONAL TEST INCLUDES BOWING THE SPHINCTEROTOME TO ENSURE THE DISTAL END RESPONDS TO HANDLE MANIPULATION. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK FUSION OMNI-TOME. IT WAS INITIALLY REPORTED THAT THE PHYSICIAN DETECTED THE DISTAL END OF CUTTING WIRE BROKEN AFTER PLUG IN WITH ELECTRICITY. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. PER COMMUNICATION RECEIVED ON (B)(6)2022: CUTTING WIRE ANCHOR DISCONNECTED AT THE PATIENT END OF DEVICE. [SUBJECT OF REPORT] THE DEVICE WAS RETURNED WITH A SECTION MISSING; HOWEVER THE INITIAL REPORTER STATED THAT A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
571173 FUSION OMNI-TOME KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS WILSON-COOK MEDICAL INC W4337874 10827002319032

Patients

Seq Age Sex Outcome Treatment
1 72 YR Male COOK METII-35-480 WIRE GUIDE| ERBE ELECTROSURGICAL GENERATOR| OLYMPUS TFTYGE 260V ENDOSCOPE