BD INSYTE¿ AUTOGUARD¿ PERIPHERAL INTRAVENOUS CATHETER
Report
- Report Number
- 9610048-2022-00072
- Event Type
- Malfunction
- Date Received
- July 22, 2022
- Date of Event
- June 22, 2022
- Report Date
- September 19, 2022
- Manufacturer
- BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
- Product Code
- FOZ
- UDI-DI
- 00382903818235
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: A PHYSICAL SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION BUT BD WAS PROVIDED WITH A PHOTO OF THE ISSUE FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED FOR THE REPORTED LOT, 132746, AND NO QUALITY ISSUES WERE FOUND DURING PRODUCTION. OUR QUALITY ENGINEER REVIEWED THE PROVIDED PHOTO AND OBSERVED THAT THE NEEDLE HAD PIERCED THROUGH THE CATHETER TUBING. THEREFORE, BASED OFF THE PROVIDED PHOTOS THE ENGINEER WAS ABLE TO VERIFY THE REPORTED DEFECT. UNFORTUNATELY, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED AS THE DEVICE APPEARED TO HAVE BEEN USED. THIS TYPE OF DEFECT CAN OCCUR DURING THE CATHETER PLACEMENT PROCESS OR DURING THE PROCESS OF BREAKING TIP ADHESION. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT WHILE USING BD INSYTE¿ AUTOGUARD¿ PERIPHERAL INTRAVENOUS CATHETER CALIBER IS TEARING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: CUSTOMER REPORTS THAT IAG 22G CALIBER IS TEARING.
IT WAS REPORTED THAT WHILE USING BD INSYTE¿ AUTOGUARD¿ PERIPHERAL INTRAVENOUS CATHETER CALIBER IS TEARING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: CUSTOMER REPORTS THAT IAG 22G CALIBER IS TEARING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 608094 | BD INSYTE¿ AUTOGUARD¿ PERIPHERAL INTRAVENOUS CATHETER | CATHETER, INTRAVASCULAR | FOZ | BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. | 381823 | 0132746 | 00382903818235 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |