FDA Adverse Event Malfunction Summary report: N

MMSI ROD, 5.5 X 45MM, TI

MDR report key: 15085913 · Received July 22, 2022

Report

Report Number
1526439-2022-01265
Event Type
Malfunction
Date Received
July 22, 2022
Date of Event
June 24, 2022
Manufacturer
DEPUY SPINE INC
Product Code
MNH
UDI-DI
10705034077314
PMA / PMN Number
K955348
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL PRODUCT CODE: NKB, KWQ, KPW AND MNI. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. INVESTIGATION SUMMARY : THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THERE WAS NO DAMAGE OR DEFECTS WITH THE MMSI ROD, 5.5 X 45MM, TI. A DIMENSIONAL INSPECTION WAS PERFORMED FOR THE MMSI ROD, 5.5 X 45MM, TI AND MET SPECIFICATIONS. A FUNCTIONAL TEST COULD NOT BE PERFORMED AS THE MATING DEVICE WAS NOT RETURNED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE MMSI ROD, 5.5 X 45MM, TI WAS FOUND TO HAVE NO DAMAGE OR DEFECTS. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 179762045, LOT NUMBER: BDO1EVN; AND NO NON-CONFORMANCES /MANUFACTURING IRREGULARITIES RELATED TO THE MALFUNCTION WERE IDENTIFIED.

Description of Event or Problem · 0

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A PLIF (L4-5) FOR TREATING SPONDYLOLISTHESIS ON (B)(6) 2022. THE 7.0X 40MM SCREW (UNK) CUT OUT DURING SCREW TIGHTENING (WIPER MOTION) WITH THE TORQUE WRENCH IN QUESTION. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT (7.0X 45MM SCREW) LESS THAN 30-MINUTE SURGICAL DELAY. HE COMMENTED THAT THE EVENT MIGHT HAVE BEEN TRIGGERED BY MALFUNCTIONING TORQUE-VALUE SETTING ON THE TORQUE WRENCH. IT WAS BECAUSE AN EXCESSIVE LOAD HAD BEEN UNEXPECTEDLY PUT ONTO THE SCREW. WHEN THE SURGEON WAS APPLYING THE REPLACEMENT, HE ALSO FELT STRONG LOAD EVEN BEFORE TORQUE CAME INTO ACTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) MMSI ROD, 5.5 X 45MM, TI. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86968 MMSI ROD, 5.5 X 45MM, TI ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION MNH DEPUY SPINE INC 179762045 BDO1EVN 10705034077314

Patients

Seq Age Sex Outcome Treatment
1 Unknown