FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGE

MDR report key: 15085217 · Received July 22, 2022

Report

Report Number
1920898-2022-00472
Event Type
Malfunction
Date Received
July 22, 2022
Date of Event
June 22, 2022
Report Date
July 28, 2022
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED (5) 3/10CC, 12.7MM, 29G SYRINGES IN AN OPEN POLY BAG FROM LOT # 1200370. CUSTOMER STATES THAT WHEN THE PLUNGER ROD WAS PUSHED, A LIQUID CAME OUT IN FIVE OF SEVEN PRODUCTS. ALL RETURNED SYRINGES WERE TESTED AND A CLEAR DROP OF MATERIAL CAME OUT OF THE CANNULA OF 4 OUT OF 5 SAMPLES. A SMALL PORTION OF THIS MATERIAL WAS REMOVED FROM THE SAMPLE AND PREPARED FOR FTIR SPECTRAL ANALYSIS. THE SPECTRAL ANALYSIS SHOWS THAT THIS MATERIAL IS MOST LIKELY SILICONE. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 1200370 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE THREE (3) NOTIFICATIONS NOTED FOR POOLING. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT-CAUSE: BASED ON (B)(4), OPERATOR FOUND EXCESS LUBE INSIDE THE BARRELS DURING IN-PROCESS INSPECTION, THE DEFECT OCCURRED DUE TO DAMAGED ADHESIVE NOZZLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT FIVE OF THE SEVEN BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGES HAD A FOREIGN MATTER COME OUT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN THE HCP PUSHED THE PLUNGER ROD, A LIQUID CAME OUT IN FIVE OF SEVEN PRODUCTS.¿

Description of Event or Problem · 0

IT WAS REPORTED THAT FIVE OF THE SEVEN BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGES HAD A FOREIGN MATTER COME OUT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN THE HCP PUSHED THE PLUNGER ROD, A LIQUID CAME OUT IN FIVE OF SEVEN PRODUCTS.¿

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
566137 BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 1200370

Patients

Seq Age Sex Outcome Treatment
1 Unknown