FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ TIP SYRINGE

MDR report key: 15077022 · Received July 21, 2022

Report

Report Number
1911916-2022-00361
Event Type
Malfunction
Date Received
July 21, 2022
Date of Event
June 28, 2022
Report Date
July 15, 2022
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903096535
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: IT WAS REPORTED THERE WAS FOREIGN MATTER. TO AID IN THE INVESTIGATION, TWO SAMPLES IN OPENED PACKAGING BLISTERS AND TWO PHOTOS WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED AND THE SYRINGE BARRELS HAVE EMBEDDED DEGRADED RESIN. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THE TWO PHOTOS WERE PROVIDED SHOW THE SAMPLES RECEIVED. THE EMBEDDED DEGRADED RESIN IN THE COMPONENT TYPICALLY OCCURS AT THE STARTUP OR INTERMITTENTLY DURING THE INJECTION MOLDING PROCESS. THE DEGRADED RESIN CAN BREAK LOOSE AND BE MOLDED INTO COMPONENTS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309653, LOT NUMBER 2090599. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THE FREQUENCY OF INSPECTIONS WERE INCREASED TO MITIGATE THE ESCAPE OF THIS DEFECT. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE HAD FOREIGN MATTER IN IT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE HAVE SOME RECENT CONCERNS REGARDING SEVERAL EACHES OF A SPECIFIC LOT OF ITEM 309653."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1654074 BD LUER-LOK¿ TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 309653 2090599 30382903096535

Patients

Seq Age Sex Outcome Treatment
1 Unknown